Skip to content

TECNIS IOL ZCT400 (REF ZCT400U120)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TECNIS TORIC 1-PIECE 6.0MM 12.0D CYL4.00 · Class 3 · Implant

All models of TECNIS IOL (1344)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05050474553750), received between Apr 3, 2017 and Dec 29, 2018 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Unintended Movement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Unintended Movement1

Patient problems

  • No Code Available1
  • No Consequences Or Impact To Patient1
  • Visual Disturbances1
  • Visual Impairment1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 29, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT EXPERIENCED ANSIOMETROPIA AFTER THE IMPLANTATION OF A MODEL ZCT400 12.0 DIOPTER TECNIS TORIC IOL (INTRAOCULAR LENS). THE LENS WAS EXPLANTED AND REPLACED WITH A MODEL ZCT225 12.5 DIOPTER LENS. PATIENT WAS DOING WELL POST EXCHANGE. IT WAS NOTED THAT THERE WERE NO INCISION ENLARGEMENT, NO SUTURE AND NO VITRECTOMY REQUIRED. VISUAL ACUITY PRE-INITIAL IMPLANT: 20/400; VISUAL ACUITY POST-INITIAL IMPLANT: 20/60-1; VISUAL ACUITY POST-REPLACEMENT:20/20. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Code Available; Visual Disturbances; Visual Impairment

    MAUDE report 8205382 ↗
  • MalfunctionApr 3, 2017Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT A TECNIS TORIC INTRAOCULAR LENS (IOL) WAS IMPLANTED IN THE RIGHT EYE OF A (B)(6) FEMALE PATIENT. THE IOL, INITIALLY PLACED AT 108 DEGREES, WAS MEASURED AT THE SIX-MONTH POST-OPERATION VISIT AT 96 DEGREES, WITH A DIFFERENCE OF 12 DEGREES. REPORTEDLY, THERE WERE NO PATIENT OR DOCTOR COMPLAINTS AND THE BCDVA (BEST-CORRECTED DISTANCE VISUAL ACUITY) WAS 20/20 ON BOTH EYES. A TECNIS TORIC ROTATION, FOR THE LENS IMPLANTED IN THE PATIENT'S LEFT EYE, WAS ALSO REPORTED. A SECOND MDR WILL BE FILED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6455388 ↗

Other LENS, INTRAOCULAR, TORIC OPTICS devices