Skip to content

TECNIS IOL ZLB00 (REF ZLB00U0080)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TECNIS MF 1- PC 6.0MM 8.0D LOW ADD US · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05050474560437), received between Mar 31, 2017 and Mar 31, 2017 — classified as 1 injury.

The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Inadequacy of Device Shape and/or Size1

Patient problems

  • No Code Available1
  • Visual Disturbances1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 31, 2017Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT A A ZLB00 8.0 DIOPTER (D) INTRAOCULAR LENS (IOL) WAS EXPLANTED DUE TO PATIENT EXPERIENCING SEVERE DISTORTED VISION FOLLOWING THE CATARACT SURGERY. THE LENS WAS REPLACED WITH THE SAME MODEL LENS WITH A 9.5 DIOPTER. THE WAS NO INCISION ENLARGEMENT OR VITRECTOMY REQUIRED. THERE WAS NO PATIENT INJURY REPORTED. THROUGH FOLLOW UP IT WAS CONFIRMED PHYSICIAN PREFORMED THE IOL EXCHANGE DUE TO A MISCALCULATION AND OVERCORRECTION OF THE IOL POWER. ADDITIONALLY, IT WAS LEARNED THAT THE PATIENT'S REFRACTION WAS +1.25 BEST CORRECTED 20/60 AFTER THE ZLB00 8.0 D IOL WAS IMPLANTED. PRIOR TO SURGERY PATIENT'S LEFT EYE (OS) WAS BEST CORRECTED 20/25 WITH -8.25 SPHERE (SPH).

    Patient problems: No Code Available; Visual Disturbances

    MAUDE report 6450446 ↗

Other LENS, MULTIFOCAL INTRAOCULAR devices

All LENS, MULTIFOCAL INTRAOCULAR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page