The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05050474603370), received between Nov 2, 2016 and Sep 28, 2018 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Unintended Movement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Unintended Movement1

Patient problems

  • No Code Available1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAD UNEXPECTED POST-OP REFRACTION AFTER IMPLANT OF A MODEL ZCT525 TECNIS TORIC IOL (INTRAOCULAR LENS) IN HIS RIGHT EYE (OD). PRE-OP REFRACTION: -7.75 -7.75 X 070; POST-OP REFRACTION: -3.00 -4.75 X 045; UCVA (UNCORRECTED VISUAL ACUITY): CFS (COUNTING FINGERS); BCVA (BEST CORRECTED VISUAL ACUITY): 20/200. THROUGH FOLLOW-UP, THE LENS WAS EXPLANTED AND REPLACED WITH A MODEL AR40E. THE PATIENT WAS SEEN ON (B)(6) 2018. VISION WAS 2100 AND PATIENT CAN ONLY COUNT FINGERS. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Code Available

    MAUDE report 7920566 ↗
  • MalfunctionNov 2, 2016Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT A TECNIS TORIC INTRAOCULAR LENS (IOL) WAS IMPLANTED IN THE RIGHT EYE OF A (B)(6) MALE PATIENT. THE IOL, INITIALLY PLACED AT 95 DEGREES, WAS MEASURED AT SIX-MONTH VISIT AT 84 DEGREES. REPORTEDLY, NEITHER THE SURGEON OR THE PATIENT HAD ANY COMPLAINTS AT THE SIX-MONTH VISIT. THE LENS REMAINED IMPLANTED AND THE BCDVA (BEST CORRECTED DISTANCE VISUAL ACUITY) WAS 20/25. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6073654 ↗