Skip to content

TECNIS IOL ZCT600 (REF ZCT600U340)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TECNIS TORIC 1-PIECE 6.MM 34.0D CYL6.00 · Class 3 · Implant

All models of TECNIS IOL (1344)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05050474604445), received between Mar 3, 2021 and Mar 3, 2021 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Visual Impairment1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 3, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WILL BE EXPLANTED AND REPLACED WITH A DIFFERENT MODEL, ZCU375, AND LOWER DIOPTER POWER (28.0) IOL. IT WAS LEARNED THAT THE PATIENT HAD EXTREME HYPEROPIA. LADAS, BARRETT, & LENSTAR INTERNAL FORMULAS PREDICTED EMMETROPIA WITH 34.5D LENS. REPORTER CONSULTED WITH AN EXPERT WHO THINKS THAT J&J LENSES MAY BE OVERPOWERED AT HIGH POWERS. IT WAS IN THIS CASE. PRE-OP VISUAL ACUITY (VA) 20/70, POST-OP 20/25, BUT -5.25 + 1.25. AT EXPLANT THERE WAS NO INCISION ENLARGEMENT, NO VITRECTOMY, AND NO SUTURES WERE REQUIRED. NO ADVERSE EVENTS. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Visual Impairment

    MAUDE report 11407992 ↗

Other LENS, INTRAOCULAR, TORIC OPTICS devices

All LENS, INTRAOCULAR, TORIC OPTICS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page