SMARTLOAD GCB00 (REF GCB0000235)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · SMARTLOAD TECNIS 1PC CLEAR MONO 23.5D · Class 3

All models of SMARTLOAD (52)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 05050474620674), received between Feb 8, 2021 and May 19, 2026 — classified as 1 injury, 12 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Contamination3
  • Physical Resistance/Sticking2
  • Adverse Event Without Identified Device or Use Problem1
  • Delivered as Unsterile Product1
  • Difficult to Fold, Unfold or Collapse1
  • Failure to Eject1
  • Failure to Fold1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • Eye Injury1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2026Contamination

    IT WAS REPORTED THAT POST IMPLANTATION OF THE PRELOADED, MONOFOCAL, INTRAOCULAR LENS (IOL) THE SURGEON OBSERVED A BLACK FIBER IN THE PATIENT'S EYE. THERE WAS NO PATIENT INJURY REPORTED AND NO INTERVENTION WAS REQUIRED. THE SURGEON NOTED THAT THE PATIENT WAS FEMALE AND INDICATED THAT THE BLACK FIBER COULD BE MAKEUP NOT PROPERLY REMOVED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25225141 ↗
  • MalfunctionDec 29, 2025Material Opacification; Material Opacification

    IT WAS REPORTED THAT, ON A COUPLE OF OCCASIONS, THE SURGEON OBSERVED THAT A LENS APPEARED ¿FURRY¿ WHEN VIEWED UNDER A MICROSCOPE DURING PROCEDURES. THIS ISSUE WAS NOT VISIBLE TO THE SCRUB NURSE BUT WAS CLEARLY SEEN BY THE SURGEON UNDER THE MICROSCOPE. THROUGH FOLLOW UP WE LEARNED THERE WERE 6 OCCASIONS OF THIS ISSUE ALL OBSERVED ON THE SAME DAY. THE SURGEON BELIEVES THE IOLS WERE NOT POLISHED AND THE ISSUE WAS ONLY LOCATED ON THE EDGE OF THE OPTICS. NO PATIENT HARM OR ADVERSE OUTCOMES WERE REPORTED. NO FURTHER INFORMATION WAS PROVIDED. THIS REPORT CAPTURES ONE EVENT, THE OTHER FIVE (5) EVENTS ARE REPORTED SEPARATELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23917924 ↗
  • MalfunctionDec 13, 2025Break; Break

    IT WAS REPORTED THAT THE PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) WAS BROKEN. NO FURTHER DETAILS WERE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23798693 ↗
  • MalfunctionMay 21, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT THE PACKAGE OF THE INTRAOCULAR LENS (IOL) WAS NOT SEALED WHEN THE BOX WAS OPENED. THROUGH FOLLOW-UP, WE LEARNED THAT THE CUSTOMER DOES NOT THINK IT WAS CAUSED BY SHIPPING BECAUSE THE OUTER BOX WAS INTACT. NO ADDITIONAL INFORMATION WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22064330 ↗
  • MalfunctionJul 22, 2024Contamination; Contamination

    IT WAS REPORTED THAT AFTER INSERTING THE PRELOADED INTRAOCULAR LENS (IOL), IT WAS DISCOVERED THAT THERE WAS A SMALL U SHAPED PIECE OF PLASTIC IN THE PATIENT'S EYE, WHICH HAD APPARENTLY COME OUT OF THE SHOOTER WITH THE LENS, POSSIBLY FROM THE CARTRIDGE. THE PATIENT IS NOT AFFECTED, THE SURGERY WENT AS PLANNED. THE PLASTIC PART WAS ASPIRATED OUT OF THE EYE, BUT IS NOT AVAILABLE FOR TESTING. THE CARTRIDGE IS AVAILABLE. NO FURTHER INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19798254 ↗
  • MalfunctionNov 22, 2023Break; Break

    IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS (IOL) WAS DEFECTIVE. NO FURTHER INFORMATION WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18196140 ↗
  • MalfunctionMar 10, 2023Contamination; Contamination

    IT WAS REPORTED THAT DURING SURGERY A PIECE OF ANOTHER INTRAOCULAR LENS (IOL), APPEARING TO BE A DETACHED HAPTIC, WAS FOUND IN THE PATIENT'S EYE. THE SURGEON REMOVED IT WITHOUT ANY ISSUES. THE IOL REMAINS IMPLANTED AND THE PATIENT HAS FULLY RECOVERED. NO FURTHER INTERVENTIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16523854 ↗
  • InjuryMar 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) WAS REMOVED. NO FURTHER INFORMATION PROVIDED.

    Patient problems: Eye Injury

    MAUDE report 16500071 ↗
  • MalfunctionJan 13, 2023Difficult to Fold, Unfold or Collapse; Physical Resistance/Sticking; Difficult to Fold, Unfold or Collapse; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE HAPTICS OF THE INTRAOCULAR LENS (IOL) WERE STUCK TO EACH OTHER WHEN PLACED WITHIN THE PATIENT'S OPERATIVE EYE. ACCOUNT INDICATED THAT THE HAPTICS WERE KISSING TIGHT. IT TOOK SEVERAL MINUTES TO MAKE THE IOL OPEN-UP. THROUGH FOLLOW UP WE LEARNED THAT ADDITIONAL MANEUVERS WERE NEEDED TO RELEASE THE HAPTICS. THE TOOL OLIVE BURATTO WAS USED. NO ADDITIONAL INFORMATION WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16162396 ↗
  • MalfunctionDec 15, 2022Break; Break

    IT WAS ORIGINALLY REPORTED THAT THE SMARTLOAD INTRAOCULAR LENS (IOL) WAS BROKEN AND IT WAS NOT IMPLANTED. THROUGH FOLLOW-UP WE LEARNED THAT THE IOL WAS IMPLANTED AND ONCE INSIDE THE EYE, THE SURGEON SAW THAT THE LENS WAS BROKEN. THE IOL WAS CUT AND REMOVED FROM THE PATIENT'S OPERATIVE EYE. ANOTHER LENS, SAME MODEL, WAS IMPLANTED WITHOUT ANY MAJOR DIFFICULTIES. THERE WAS NO PATIENT INJURY REPORTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15992884 ↗
  • MalfunctionSep 28, 2022Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE HAPTICS OF THE IOL (INTRAOCULAR LENS) WERE STUCK TO EACH OTHER. IT HAS TAKEN SEVERAL MINUTES TO SEPARATE THE HAPTICS. THROUGH FOLLOW UP WE LEARNED THAT THERE WAS NO HARM TO THE PATIENT AFTER THE IMPLANT OF THE LENS. THE IOL REQUIRED ADDITIONAL MANEUVERS TO POSITION THE LENS BECAUSE THE CUSTOMER HAD TO OPEN THE HAPTICS MANUALLY WITH THE HELP OF A SURGICAL INSTRUMENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15499572 ↗
  • MalfunctionMay 17, 2021Break; Break

    IT WAS REPORTED THAT THE CUSTOMER SENT IN A FAX INDICATING THE LENS WAS UNSTERILE / DEFECTIVE. NO FURTHER DETAILS WERE GIVEN. PRODUCT INVESTIGATION WAS COMPLETED WITHOUT SAMPLE RETURN. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11833689 ↗
  • MalfunctionFeb 8, 2021Failure to Fold; Failure to Eject; Failure to Fold; Failure to Eject

    IT WAS REPORTED THAT THE TRAILING HAPTIC WAS STRAIGHT BEHIND THE LENS AND IT WAS STUCK BETWEEN CARTRIDGE TIP AND ROD. THE HAPTIC WAS MANIPULATED GENTLY AND RELEASED WITH SINSKEY HOOK INSTRUMENT. THE HAPTIC WAS PLACED IN THE BAG AND LENS CENTERED. NO EXTRA INTERVENTIONS NEEDED. THE LENS WAS FULLY INSERTED, THE PATIENT WAS DISCHARGED ACCORDING NORMAL PROTOCOL. NO POST OP. CHECK NEEDED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11298826 ↗

Other intraocular lens devices

All intraocular lens devices →