Skip to content

TECNIS SIMPLICITY DCB00 (REF DCB0000050)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TECNIS SMPLCTY TECNIS 1PC CLR MONO 5.0D · Class 3 · Implant

All models of TECNIS SIMPLICITY (112)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05050474635784), received between Jan 26, 2021 and Nov 3, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Mechanical Jam2
  • Scratched Material2
  • Crack1
  • Device Damaged by Another Device1
  • Device Difficult to Setup or Prepare1
  • Difficult to Fold, Unfold or Collapse1
  • Ejection Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 3, 2025Device Difficult to Setup or Prepare; Scratched Material; Device Difficult to Setup or Prepare; Scratched Material

    IT WAS REPORTED THAT ISSUES OCCURRED WHILE SLIDING THE PLUNGER TO LOAD THE PRELOADED INTRAOCULAR LENS (IOL). DESPITE APPLYING FORCE, THE LENS WAS SUCCESSFULLY LOADED AND EJECTED; HOWEVER, IT COULD NOT BE INSERTED AS IT BECAME SCRATCHED. THROUGH FOLLOW-UP, WE LEARNED THAT THE SURGEON FORCED THE LENS DURING HANDLING ON THE OVER-THE-COUNTER TABLE, WHICH CAUSED IT TO COME OUT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23459502 ↗
  • MalfunctionJun 16, 2025Break; Mechanical Problem; Break; Mechanical Problem

    IT WAS REPORTED THAT DURING THE IMPLANTATION OF THE PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL), THE LENS APPEARED WRINKLED AND MARKED, LEADING TO ITS DISPOSAL. THROUGH FOLLOW-UP, WE LEARNED, THAT THE EVENT OCCURRED ON MAY 15TH, 2025 AND THE LENS RUPTURED DUE TO AN ERROR IN THE PRELOADING SYSTEM, THE PLUNGER PASSED OVER THE LENS. THE PATIENT WAS IMPLANTED WITH A REPLACEMENT DEVICE. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22216417 ↗
  • MalfunctionMar 7, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    IT WAS REPORTED THAT A PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) COULD NOT UNFOLD. THERE WAS PATIENT CONTACT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21554682 ↗
  • MalfunctionJun 21, 2024Break; Mechanical Jam; Break; Mechanical Jam

    IT WAS REPORTED THAT THE PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) WAS DEFECTIVE. THROUGH FOLLOW-UP, WE LEARNED THAT THE IOL GOT STUCK INSIDE THE INJECTOR AND THE HAPTIC WAS TORN. THE TIP OF THE CARTRIDGE WAS IN CONTACT WITH THE PATIENT'S EYE. NO ADDITIONAL MANEUVERS WERE PERFORMED. A REPLACEMENT IOL WAS IMPLANTED. NO FURTHER DETAILS WERE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19581090 ↗
  • MalfunctionSep 14, 2023Crack; Mechanical Jam; Ejection Problem; Crack; Mechanical Jam; Ejection Problem

    IT IS REPORTED THAT A PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) TRAILING HAPTIC STUCK IN THE INJECTOR DURING IMPLANTATION OR APPLICATION. SURGEON WAS ABLE TO FULLY INSERT AND USE THE LENS (IMPLANTED). NO PATIENT INJURY OR MEDICAL AND SURGICAL INTERVENTION IS REQUIRED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17746268 ↗
  • MalfunctionJan 26, 2021Device Damaged by Another Device; Scratched Material; Device Damaged by Another Device; Scratched Material

    IT WAS REPORTED THAT A DCB00 PRE-LOADED DEVICE INJECTOR MALFUNCTION, SCRATCHED LENS. DURING HANDLING BUT PRIOR TO INSERTION. NO PATIENT CONTACT. NO ADVERSE EFFECT, INJURY OR SURGICAL INTERVENTION REQUIRED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11231902 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page