TECNIS SIMPLICITY DCB00 (REF DCB0000280)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TECNIS SMPLCTY TECNIS 1PC CLR MONO 28.0D · Class 3 · Implant
All models of TECNIS SIMPLICITY (112)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05050474636248), received between Jun 30, 2023 and Dec 8, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Crack2
- Failure to Fold1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 8, 2025Break; Break
IT WAS REPORTED THAT WHEN INTRAOCULAR LENS (IOL) WAS IMPLANTED INTO THE EYE AND WHILE DOCTOR WAS ADJUSTING AND POSITIONING THE HAPTIC, THE HAPTIC BROKE. IOL WAS REMOVED AND REPLACED. SURGERY COMPLETED SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23751111 ↗MalfunctionAug 22, 2025Crack; Crack
IT WAS REPORTED THAT DURING IMPLANTATION OF THE PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) THE IOL WAS TRACKING ON THE SIDE OF THE TIP. THE CARTRIDGE TIP WAS IN CONTACT WITH THE EYE BUT NO PATIENT INJURY WAS REPORTED. NO MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED. NO CONCERNS WERE RAISED REGARDING PATIENT OUTCOME. NO FURTHER INFORMATION AS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22877833 ↗MalfunctionJan 15, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) WAS FULLY INSERTED WHEN IT WAS NOTICED THAT THE HAPTIC WAS CAUGHT IN THE INJECTOR AND BECAME DISLOCATED. THE IOL WAS REMOVED WITH FORCEPS, AND THE PROCEDURE WAS COMPLETED WITH A BACKUP LENS. THERE WAS NO PATIENT INJURY REPORTED. NO ADDITIONAL INFORMATION WAS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21162023 ↗MalfunctionDec 21, 2023Break; Failure to Fold; Break; Failure to Fold
IT WAS REPORTED BY THE SURGEON THAT THE LEADING HAPTIC OF THE INTRAOCULAR LENS DID NOT COME OUT OF THE PRELOADED CARTRIDGE CORRECTLY, AND WAS DEFECTIVE. LENS WAS PARTIALLY INSERTED IN THE PATIENTS LEFT EYE. A NEW LENS HAD TO BE OPENED AND IMPLANTED. THE ISSUE WAS IDENTIFIED DURING IMPLANTATION. DIRECTIONS FOR USE WERE FOLLOWED. THERE WAS A DELAY IN THE PROCEDURE. PATIENT OUTCOME WAS UNKNOWN. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18372660 ↗MalfunctionNov 20, 2023Crack; Crack
IT WAS REPORTED THE DCB00 MODEL INTRAOCULAR LENS (IOL) INSERTER CRACKED WHILE IN THE PATIENT'S OPERATIVE EYE. WHEN THE DOCTOR WENT TO IMPLANT THE IOL, HE SAW THAT THE TIP WAS STRESSED, LOOKED LIKE IT WAS GOING TO CRACK. THE DOCTOR STATED HE HAS SEEN THIS BEFORE, SO HE DECIDED TO NOT FULLY ADVANCE THE LENS AND USE A NEW ONE. HE FULLY THE ADVANCED THE DEFECTED INSERTER OUTSIDE THE PATIENTS EYE AND THE TIP CRACKED. THE DOCTOR DID THIS JUST TO CONFIRM WHAT HE THOUGHT WAS GOING TO HAPPEN. THERE WAS NO ADDITIONAL INTERVENTION REQUIRED. PATIENT STATUS POST-PROCEDURE WAS REPORTED AS DOING WELL. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18179066 ↗MalfunctionJun 30, 2023Break; Break
IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS (IOL) BROKE DURING THE IMPLANTATION. THE LENS CAME OUT WITH A BROKEN PROXIMAL HAPTIC FROM THE INJECTOR. THE IOL WAS REMOVED DURING THE SAME PROCEDURE. THROUGH FOLLOW UP WE LEARNED THAT A NON-JOHNSON AND JOHNSON LENS WAS IMPLANTED SUCCESSFULLY. NO ADDITIONAL INFORMATION WAS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17243873 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page