TECNIS SIMPLICITY DCB00 (REF DCB0000320)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TECNIS SMPLCTY TECNIS 1PC CLR MONO 32.0D · Class 3 · Implant
All models of TECNIS SIMPLICITY (112)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05050474636323), received between May 2, 2023 and May 16, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Mechanical Jam2
- Break1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 16, 2023Material Fragmentation; Mechanical Jam; Material Fragmentation; Mechanical Jam
IT WAS REPORTED THE PRE-LOADED DCB00 MODEL INTRAOCULAR LENS (IOL) TRAILING HAPTIC WAS WEDGED BETWEEN THE CARTRIDGE AND PADDLE, AND WAS UNABLE TO RELEASE. ULTIMATELY, THE HAPTIC BROKE OFF AND THE IOL HAD TO BE EXPLANTED. THERE WAS PATIENT CONTACT HOWEVER, EXTENT OF CONTACT IS UNKNOWN. THE SAME PROCEDURE WAS COMPLETED SUCCESSFULLY USING J&J BACKUP LENS WITH THE SAME MODEL AND DIOPTER SIZE. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16943372 ↗MalfunctionMay 2, 2023Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT AT THE TIME OF IMPLANTATION OF THE PRELOADED INTRAOCULAR LENS (IOL), WHEN SLIDING THE IMPLANT SYSTEM PLUNGER, THE LENS GOT STUCK ON THE TIP OF THE CARTRIDGE AND WHEN TRYING TO PASS THE LENS, IT WAS DAMAGED. ONLY THE CARTRIDGE TIP MADE CONTACT WITH THE EYE. THERE WAS NO BACK-UP LENS OF THE SAME DEGREE, SO THE PATIENT HAD TO REMAIN APHAKIC (WITHOUT LENS) UNTIL THE ARRIVAL OF THE NEW LENS. THE PATIENT REMAINED APHAKIC FOR TWO DAYS AND IT IS NOTED THAT THE LENS PLACEMENT SURGERY IS NOT YET PERFORMED AND PATIENT IS TEMPORARILY IMPAIRED. NO INCISION ENLARGEMENT OR VITRECTOMY REQUIRED. IT IS NOTED THAT THE END-USER SOUGHT MEDICAL CARE. STANDARD POST-OP MEDICATION WAS PRESCRIBED. INSTRUCTIONS FOR USE WERE FOLLOWED. THROUGH FOLLOW-UP, IT WAS LEARNED THAT ANOTHER JOHNSON AND JOHNSON IOL (ZCB, 32.0D) WAS IMPLANTED IN THE RIGHT EYE DURING A NEW STANDARD SURGERY ON (B)(6) 2023, WHICH INVOLVED ANESTHESIA, SEDATION, INCISION (MAIN AND PARACENTESIS). THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. THE PATIENT OUTCOME IS "GOOD EVOLUTION". THE PATIENT WAS DISCHARGED UNEVENTFULLY AND THERE ARE NO PLANNED INTERVENTIONS. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16856107 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page