TECNIS IOL ZCU225 (REF ZCU225U190)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TECNIS TORIC II 1-PC 19.0D CYL2.25 · Class 3 · Implant
All models of TECNIS IOL (1344)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05050474658950), received between Feb 25, 2021 and Jul 11, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Unstable” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Unstable1
Patient problems
- Eye Injury1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE ZCU225 MODEL INTRAOCULAR LENS (IOL) WAS IMPLANTED IN THE PATIENT¿S OCULAR SINISTER (LEFT EYE). DUE TO PATIENT PREFERENCE TO CHANGE FROM A TORIC TO A MULTI-FOCAL LENS, THE IOL WAS EXPLANTED IN A SECONDARY SURGICAL PROCEDURE AND REPLACED WITH A DRT225 19.0 IOL. THERE WAS NO MEDICAL INTERVENTION, NO PATIENT INJURY, AND NO SURGICAL INTERVENTION DURING THE LENS EXCHANGE PROCEDURE. PATIENT PROGNOSIS POST LENS EXCHANGE PROCEDURE IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Eye Injury
MAUDE report 22487654 ↗MalfunctionFeb 25, 2021Unstable; Unstable
IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WAS IMPLANTED IN A STUDY SUBJECT AND THE OPERATIVE IOL AXIS ESTIMATE RIGHT EYE (OD) WAS 114 DEGREES. BUT AT 3-MONTH STUDY POSTOPERATIVE IOL AXIS ESTIMATE OD: 102 DEGREES. UNCORRECTED VISUAL ACUITY OD AT THE 3-MONTH POSTOPERATIVE VISIT: 20/12.5. MANIFEST REFRACTION: PLANO. BEST-CORRECTED DISTANCE VISUAL ACUITY: 20/12.5. SUBJECT REPORTED NO VISUAL SYMPTOMS AT THIS VISIT. NO SURGICAL INTERVENTION IS PLANNED AT THIS TIME. THE COMPLAINT WAS BASED UPON A SUBJECTIVE ESTIMATE MADE BY THE INVESTIGATOR FROM A SLIT-LAMP MEASUREMENT. FOR THIS STUDY, WE HAVE ALSO BEEN DOING INDEPENDENT ANALYSIS OF THE IOL PHOTOS TAKEN AT EACH VISIT. THESE ARE OBJECTIVE MEASUREMENTS OF THE ACTUAL IOL AXIS. BASED ON THIS OBJECTIVE MEASUREMENT, THE ACTUAL AMOUNT OF IOL ROTATION FOR THE SUBJECT BELOW IS LESS THAN 01 DEGREE. THERE IS STILL NO SURGICAL INTERVENTION PLANNED. THE SUBJECT HAS COMPLETED THE STUDY. SUBJECT: (B)(6). VISIT SCHEDULE: PO 3 MONTH. PAGE: ANALYST 1 REVIEW. STUDY EYE: OD. DELTA ANGLE FROM BASE: 0.075. SUBJECT: (B)(6). VISIT SCHEDULE: PO 3 MONTH. PAGE: ANALYST 2 REVIEW. STUDY EYE: OD. DELTA ANGLE FROM BASE: 0.659.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11376938 ↗
Other LENS, INTRAOCULAR, TORIC OPTICS devices
- TECNIS IOL ZCT150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS1 recall7 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS43 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS36 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS32 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS30 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS29 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS21 reports
- TECNIS IOL DIU150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page