TECNIS IOL ZCU225 (REF ZCU225U215)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TECNIS TORIC II 1-PC 21.5D CYL2.25 · Class 3 · Implant
All models of TECNIS IOL (1344)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05050474659001), received between Jun 23, 2020 and Nov 17, 2021 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Malposition of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Malposition of Device1
- Unintended Movement1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Capsular Bag Tear1
- Edema1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryNov 17, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) THAT WAS FULLY INSERTED IN THE PATIENT'S LEFT EYE WAS REMOVED IN THE SAME PROCEDURE DUE TO A TEAR IN THE CAPSULE. THERE WAS NO PATIENT INJURY. A VITRECTOMY, AN INCISION ENLARGEMENT, AND SUTURES WERE REQUIRED. ANOTHER JOHNSON & JOHNSON LENS (DIFFERENT MODEL AND LOWER DIOPTER) WAS IMPLANTED AS A REPLACEMENT. THE PATIENT IS DOING WELL POST-OPERATIVELY. NO FURTHER INFORMATION PROVIDED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Capsular Bag Tear
MAUDE report 12833023 ↗MalfunctionDec 21, 2020Unintended Movement; Unintended Movement
REPORT OF THE STUDY PATIENT'S OPERATIVE VISIT STATED THAT THE ACTUAL INTRAOCULAR LENS (IOL) AXIS PLACEMENT WAS 176 DEGREES FOR THE LEFT EYE (OS). THE PATIENT'S SLIT LAMP AXIS MEASUREMENT OS AT 1 DAY POST-OPERATIVE VISIT WAS RECORDED AS 006 DEGREES. THIS IS A 10 DEGREE ROTATION OF THE IOL AS COMPARED TO THE AXIS PLACEMENT. THE PATIENT'S 1 WEEK OS, POST-OPERATIVE VISIT SLIT LAMP AXIS MEASUREMENT WAS RECORDED AS 006 DEGREES AGAIN. UNCORRECTED VISUAL ACUITY OS AT THE 1-DAY POSTOPERATIVE VISIT: 20/16. SUBJECT REPORTED NO VISUAL SYMPTOMS AT THIS VISIT. NO SURGICAL INTERVENTION IS PLANNED AT THIS TIME. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11049520 ↗InjuryJun 23, 2020Malposition of Device; Malposition of Device
IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WAS ROTATED TWICE, THEREFORE, THE PATIENTS EYE HAD SOME SWELLING. THROUGH FOLLOW-UP, IT WAS LEARNED DURING THE INITIAL PROCEDURE, THE SURGEON COULD NOT ALIGN THE DOTS OF THE LENS WHILE IN THE EYE, THEREFORE, THE SURGEON HAD TO ROTATE THE LENS. THE LENS WAS ROTATED AGAIN, BECAUSE THE MANIPULATION OF THE LENS IN THE EYE AFFECTED THE VISUAL ACUITY. THE LENS REMAINS IMPLANTED. THE PATIENT OUTCOME WAS UNKNOWN AS THE SURGEON HAD NOT SEEN THE PATIENT FOR A FOLLOW-UP VISIT YET. NO ADDITIONAL INFORMATION WAS PROVIDED TO JOHNSON & JOHNSON SURGICAL VISION.
Patient problems: Edema
MAUDE report 10182912 ↗
Other LENS, INTRAOCULAR, TORIC OPTICS devices
- TECNIS IOL ZCT150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS1 recall7 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS43 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS36 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS32 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS30 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS29 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS21 reports
- TECNIS IOL DIU150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page