TECNIS IOL ZCU225 (REF ZCU225U240)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TECNIS TORIC II 1-PC 24.0D CYL2.25 · Class 3 · Implant
All models of TECNIS IOL (1344)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05050474659056), received between Feb 25, 2021 and Jul 9, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Application Program Problem: Dose Calculation Error” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Application Program Problem: Dose Calculation Error1
- Device Contaminated During Manufacture or Shipping1
- Labelling, Instructions for Use or Training Problem1
Patient problems
- Visual Impairment2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Visual Disturbances1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AN EXPLANT OCCURRED DUE TO A REFRACTIVE SURPRISE. THERE WAS NO PATIENT INJURY REPORTED. THE LENS WAS REMOVED DURING A SECONDARY SURGICAL PROCEDURE AND REPLACED WITH THE SAME MODEL BUT A HIGHER DIOPTER INTRAOCULAR LENS. ADDITIONAL INFORMATION REVEALED THAT THERE WAS A CALCULATION ISSUE, AS BOTH THE PENTACAM AND ORA SUGGESTED THE USE OF THIS LENS, BUT THE PATIENT ENDED UP WITH A SIGNIFICANT REFRACTIVE ERROR AFTERWARD. ON (B)(6) 2025, IT WAS NOTED THAT THE PATIENT¿S VISION WAS NOT GOOD. HOWEVER, A ONE-WEEK REFRACTION CHECK REVEALED THAT THE PATIENT STILL HAD A CONSIDERABLE AMOUNT OF ASTIGMATISM THAT NEEDED CORRECTION. THE PATIENT IS DOING BETTER NOW. THERE WAS SOME CORNEAL EDEMA FROM THE SECOND PROCEDURE, BUT IT IS CLEARING UP. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Visual Impairment
MAUDE report 22457938 ↗InjuryMay 25, 2021Application Program Problem: Dose Calculation Error; Labelling, Instructions for Use or Training Problem; Application Program Problem: Dose Calculation Error; Labelling, Instructions for Use or Training Problem
IT WAS REPORTED THAT A TECNIS TORIC INTRAOCULAR LENS (IOL) IMPLANTED INTO THE PATIENT'S RIGHT (OD) EYE. THE LENS WAS ORIENTED AT 110°. THE PREDICTED REFRACTION WAS A -.50+ .07 AT 21°. THE SURGEON RAN THE CALCULATION TWICE. HER REFRACTION TODAY IS +1.00+2.00X 010. THE SURGEON REPEATED HER IOL MASTER AND TOPOGRAPHY TODAY AND ONCE AGAIN SHE IS SHOWING 1.95D CYLINDER @100 (HER PREV IOLM SHOWED 1.84D@106). HER CORNEAL TOPOGRAPHY IS CONSISTENT WITH THE READINGS THAT ARE GETTING ON HER IOL MASTER. THROUGH FOLLOW-UP, IT WAS LEARNED THAT THE LENS HAS EXPLANTED FROM THE PATIENT'S EYE. THE SURGEON WOULD LIKE TO KNOW IF IOL WAS MISLABELED OR IF IT WAS A MISTAKE ON THEIR END. NO FURTHER INFORMATION PROVIDED.
Patient problems: Visual Impairment
MAUDE report 11879384 ↗InjuryFeb 25, 2021Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
THE CUSTOMER REPORTED THAT A PATIENT COMPLAINED OF GLARE SYMPTOMS AT HIS ONE-MONTH POST-OPERATIVE VISIT AFTER THE IMPLANTATION OF A TORIC INTRAOCULAR LENS (IOL). THE SURGEON DILATED THE PATIENT'S EYE AND SAW A RECTANGULAR FOREIGN BODY BETWEEN THE IOL AND POSTERIOR CAPSULE. THE SURGEON HAS SEEN SMALL CORTICAL/CAPSULAR FRAGMENTS TRAPPED IN THE CAPSULE POST-OPERATIVELY WHICH DISAPPEARED WITH YAG CAPSULOTOMY, SO HE PROCEEDED WITH THE LASER, BUT THE FOREIGN BODY IS EXTREMELY FIRMLY ADHERENT, AND DOESN'T MOVE AT ALL WITH THE LASER. IT ALMOST LOOKS LIKE A PIECE OF PLASTIC OR CELLOPHANE STUCK ONTO THE POSTERIOR SURFACE OF THE IOL. THE LENS REMAINS IMPLANTED AND NO PATIENT INJURY REPORTED. NO FURTHER INFORMATION PROVIDED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Visual Disturbances
MAUDE report 11383196 ↗
Other LENS, INTRAOCULAR, TORIC OPTICS devices
- TECNIS IOL ZCT150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS1 recall7 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS43 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS36 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS32 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS30 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS29 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS21 reports
- TECNIS IOL DIU150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page