TECNIS IOL ZCU225 (REF ZCU225U240)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TECNIS TORIC II 1-PC 24.0D CYL2.25 · Class 3 · Implant

All models of TECNIS IOL (1344)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05050474659056), received between Feb 25, 2021 and Jul 9, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Application Program Problem: Dose Calculation Error” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Application Program Problem: Dose Calculation Error1
  • Device Contaminated During Manufacture or Shipping1
  • Labelling, Instructions for Use or Training Problem1

Patient problems

  • Visual Impairment2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Visual Disturbances1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AN EXPLANT OCCURRED DUE TO A REFRACTIVE SURPRISE. THERE WAS NO PATIENT INJURY REPORTED. THE LENS WAS REMOVED DURING A SECONDARY SURGICAL PROCEDURE AND REPLACED WITH THE SAME MODEL BUT A HIGHER DIOPTER INTRAOCULAR LENS. ADDITIONAL INFORMATION REVEALED THAT THERE WAS A CALCULATION ISSUE, AS BOTH THE PENTACAM AND ORA SUGGESTED THE USE OF THIS LENS, BUT THE PATIENT ENDED UP WITH A SIGNIFICANT REFRACTIVE ERROR AFTERWARD. ON (B)(6) 2025, IT WAS NOTED THAT THE PATIENT¿S VISION WAS NOT GOOD. HOWEVER, A ONE-WEEK REFRACTION CHECK REVEALED THAT THE PATIENT STILL HAD A CONSIDERABLE AMOUNT OF ASTIGMATISM THAT NEEDED CORRECTION. THE PATIENT IS DOING BETTER NOW. THERE WAS SOME CORNEAL EDEMA FROM THE SECOND PROCEDURE, BUT IT IS CLEARING UP. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Visual Impairment

    MAUDE report 22457938 ↗
  • InjuryMay 25, 2021Application Program Problem: Dose Calculation Error; Labelling, Instructions for Use or Training Problem; Application Program Problem: Dose Calculation Error; Labelling, Instructions for Use or Training Problem

    IT WAS REPORTED THAT A TECNIS TORIC INTRAOCULAR LENS (IOL) IMPLANTED INTO THE PATIENT'S RIGHT (OD) EYE. THE LENS WAS ORIENTED AT 110°. THE PREDICTED REFRACTION WAS A -.50+ .07 AT 21°. THE SURGEON RAN THE CALCULATION TWICE. HER REFRACTION TODAY IS +1.00+2.00X 010. THE SURGEON REPEATED HER IOL MASTER AND TOPOGRAPHY TODAY AND ONCE AGAIN SHE IS SHOWING 1.95D CYLINDER @100 (HER PREV IOLM SHOWED 1.84D@106). HER CORNEAL TOPOGRAPHY IS CONSISTENT WITH THE READINGS THAT ARE GETTING ON HER IOL MASTER. THROUGH FOLLOW-UP, IT WAS LEARNED THAT THE LENS HAS EXPLANTED FROM THE PATIENT'S EYE. THE SURGEON WOULD LIKE TO KNOW IF IOL WAS MISLABELED OR IF IT WAS A MISTAKE ON THEIR END. NO FURTHER INFORMATION PROVIDED.

    Patient problems: Visual Impairment

    MAUDE report 11879384 ↗
  • InjuryFeb 25, 2021Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    THE CUSTOMER REPORTED THAT A PATIENT COMPLAINED OF GLARE SYMPTOMS AT HIS ONE-MONTH POST-OPERATIVE VISIT AFTER THE IMPLANTATION OF A TORIC INTRAOCULAR LENS (IOL). THE SURGEON DILATED THE PATIENT'S EYE AND SAW A RECTANGULAR FOREIGN BODY BETWEEN THE IOL AND POSTERIOR CAPSULE. THE SURGEON HAS SEEN SMALL CORTICAL/CAPSULAR FRAGMENTS TRAPPED IN THE CAPSULE POST-OPERATIVELY WHICH DISAPPEARED WITH YAG CAPSULOTOMY, SO HE PROCEEDED WITH THE LASER, BUT THE FOREIGN BODY IS EXTREMELY FIRMLY ADHERENT, AND DOESN'T MOVE AT ALL WITH THE LASER. IT ALMOST LOOKS LIKE A PIECE OF PLASTIC OR CELLOPHANE STUCK ONTO THE POSTERIOR SURFACE OF THE IOL. THE LENS REMAINS IMPLANTED AND NO PATIENT INJURY REPORTED. NO FURTHER INFORMATION PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Visual Disturbances

    MAUDE report 11383196 ↗