TECNIS IOL ZCU300 (REF ZCU300U150)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TECNIS TORIC II 1-PC 15.0D CYL3.00 · Class 3 · Implant
All models of TECNIS IOL (1344)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05050474660052), received between Jan 7, 2021 and Feb 6, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Failure to Fold” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Failure to Fold1
- Use of Device Problem1
Patient problems
- Capsular Bag Tear1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 6, 2021Failure to Fold; Use of Device Problem; Failure to Fold; Use of Device Problem
IT WAS INDICATED THAT ALL SCRUB TECHS WERE OUT AND THE INTRAOCULAR LENS (IOL) MAY HAVE BEEN IMPROPERLY LOADED AS THE LEADING HAPTIC DID NOT FOLD PROPERLY. THE LENS WAS LOADED WITH THE FRONT HAPTIC SLIGHTLY STRAIGHT. AS THE LENS WAS INJECTED INTO THE PATIENT¿S RIGHT EYE, THE FRONT HAPTIC CAUGHT THE POSTERIOR CAPSULE AND CREATED A TEAR. THERE WAS PATIENT ZONULES ISSUE AND HAPTICS WOULD NOT SIT PROPERLY. THE LENS WAS IMPLANTED, THEN WAS CUT AND REMOVED FROM THE EYE DURING THE SAME PROCEDURE. IT WAS INDICATED THAT A VITRECTOMY WAS PERFORMED, THE SITE WAS SLIGHTLY ENLARGED TO ACCOMMODATE THE LENS AND SUTURES WERE USED TO CLOSE IT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING A NON-JOHNSON AND JOHNSON 3-PIECE LENS. THE PATIENT WAS REPORTED TO BE FINE AND NO ISSUES WERE REPORTED UPON DISCHARGE. THERE WAS NO PATIENT INJURY REPORTED. IT WAS NOTED THAT THE LENS WAS DISCARDED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Capsular Bag Tear
MAUDE report 11290203 ↗InjuryJan 7, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE TECNIS TORIC INTRAOCULAR LENS (IOL) HAPTICS WOULD NOT SIT PROPERLY UPON INSERTION BECAUSE THE PATIENT HAD ZONULES ISSUE. UNPLANNED VITRECTOMY ALONG WITH SUTURE WERE USED DURING THE PROCEDURE. THE REPLACEMENT LENS IS A NON-JOHNSON AND JOHNSON LENS. THE CUSTOMER ALSO INDICATED THAT THE LENS WAS DISCARDED. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11135355 ↗
Other LENS, INTRAOCULAR, TORIC OPTICS devices
- TECNIS IOL ZCT150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS1 recall7 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS43 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS36 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS32 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS30 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS29 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS21 reports
- TECNIS IOL DIU150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.