Skip to content

TECNIS IOL ZLU225 (REF ZLU225U185)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TECNIS TORIC LOW ADD 1-PC 18.5D CYL2.25 · Class 3 · Implant

All models of TECNIS IOL (1344)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05050474689923), received between Aug 7, 2023 and Aug 7, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2023Break; Break

    IT WAS REPORTED THE ZLU225 MODEL INTRAOCULAR LENS (IOL) FRONT HAPTIC WAS INCORRECTLY POSITIONED AS THE SURGEON STARTED TO PLACE THE LENS. THE SURGEON STOPPED AND PULLED AWAY FROM THE PATIENT, LENS WAS STILL IN CARTRIDGE, AND THERE WAS PATIENT CONTACT. IT WAS NOTED THE CARTRIDGE WAS SPLIT AND LEADING HAPTIC HAD GONE THROUGH THE CARTRIDGE. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS RECEIVED CONFIRMING THE 1MTEC30 CARTRIDGE TIP WAS CRACKED AND THE HAPTIC WAS ALSO DAMAGED. THE PROCEDURE WAS COMPLETED USING ANOTHER IOL SAME MODEL AND DIOPTER. THERE WAS NO PATIENT INJURY, NO MEDICAL INTERVENTION, AND NO SURGICAL INTERVENTION. PATIENT STATUS POST-PROCEDURE WAS REPORTED AS DOING WELL. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT CAPTURED THE HAPTIC DAMAGE ISSUE. A SEPARATE REPORT CAPTURES THE CARTRIDGE TIP DAMAGE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17477292 ↗

Other Lens, multifocal intraocular devices

All Lens, multifocal intraocular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page