Skip to content

TECNIS IOL ZLU375 (REF ZLU375U090)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TECNIS TORIC LOW ADD 1-PC 09.0D CYL3.75 · Class 3 · Implant

All models of TECNIS IOL (1344)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05050474690912), received between Dec 13, 2024 and Dec 13, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Unstable” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unstable1

Patient problems

  • Discomfort1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2024Unstable; Unstable

    IT WAS REPORTED THAT DOCTOR HAS ISSUES WITH A TORIC LENSES THAT HAS ROTATED MULTIPLE TIMES IN A PATIENT. THE INITIAL INTRAOCULAR LENS (IOL) ROTATED AND WAS EXPLANTED. THEN IT WAS ALSO REPORTED THAT THE TORIC REPLACEMENT IOL WAS IMPLANTED AT 98 DEGREES, BUT THIS ONE ALSO ROTATED A FEW DAYS AFTER IT WAS IMPLANTED IN THE PATIENT'S LEFT EYE. DAY #1 POST OP, -1.00+2.00X130 20/25, LENS AT 91 DEGREES. DAY #8 POST OP, -2.00+4.25X110 20/25 LENS AT 53 DEGREES. THEN LENS ROTATED FROM 53 DEGREES TO 95 DEGREES. IT WAS NOTED THAT AT DAY # 1 POST OP LENS ROTATION -.50+1.00X74 20/20 LENS AT 95 DEGREES. THEN AT DAY # 8 POST OP ROTATION OF LENS -100+2.00X115 20/20 LENS AT 68 DEGREES. THERE WAS SECOND LENS ROTATION AND REPOSITIONED ON 12/10/2024 LENS ROTATED FROM 68 DEGREES TO 95 DEGREES, POST OP TOMORROW TO CHECK ON STABILITY. IOL REMAINS IMPLANTED. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR THE REPLACEMENT INTRAOCULAR LENS THAT WAS REPOSITIONED ROTATIONALLY. A SEPARATE REPORT IS BEING SUBMITTED FOR THE INITIAL IOL THAT ALSO ROTATED THEN EXPLANTED.

    Patient problems: Discomfort

    MAUDE report 20926067 ↗

Other Lens, multifocal intraocular devices

All Lens, multifocal intraocular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page