Skip to content

TECNIS IOL ZLU375 (REF ZLU375U100)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TECNIS TORIC LOW ADD 1-PC 10.0D CYL3.75 · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05050474690936), received between Dec 13, 2024 and Dec 13, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Unstable” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unstable1

Patient problems

  • Visual Impairment1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 13, 2024Unstable; Unstable

    IT WAS REPORTED THAT A TORIC INTRAOCULAR LENS (IOL) WAS IMPLANTED IN THE PATIENT'S LEFT EYE AT 93 DEGREES. AT ONE DAY POST-OP, -1.00+1.50X120 20/25 AND LENS AXIS 89 DEGREES. ON DAY # 8 POST OP, -2.50 +3.50 X119 20/20, LENS AT 63 DEGREES, (SE) SPHERICAL EQUIVALENT -.75, DECIDE TO DO A LENS EXCHANGE TO ACHIEVE SE (SPHERICAL EQUIVALENT) OF ZERO. SUSPECT IOL WAS EXPLANTED ON (B)(6) 2024. THIS WAS REPLACED WITH THE SAME MODEL LENS 1.0D POWER SIZE LOWER THAN THE ORIGINAL IOL. THE REPLACEMENT IOL WAS IMPLANTED AT 98 DEGREES. SURGERY WAS COMPLETED SUCCESSFULLY. THERE WAS NO CAPSULAR TEAR, NO INCISION ENLARGEMENT, NO VITRECTOMY, NO SUTURES AND NO DIFFERENT MEDICATIONS. WHEN THE LENS IS IN THE CORRECT POSITION, THE PATIENT DOES WELL AS SHE WAS. HOWEVER, A FEW DAYS LATER REPLACEMENT LENS ROTATED AGAIN. THIS REPORT IS FOR THE INITIAL INTRAOCULAR LENS THAT WAS EXPLANTED. A SEPARATE REPORT IS BEING SUBMITTED FOR THE REPLACEMENT IOL THAT ALSO ROTATED.

    Patient problems: Visual Impairment

    MAUDE report 20929594 ↗

Other Lens, multifocal intraocular devices

All Lens, multifocal intraocular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page