TECNIS SIMPLICITY DIB00 (REF DIB00U0070)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TECNIS SIMPLICITY TECNIS EYHANCE U 07.0D · Class 3 · Implant
All models of TECNIS SIMPLICITY (112)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05050474731493), received between May 26, 2022 and Aug 31, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Mechanical Jam2
- Mechanical Problem2
- Adverse Event Without Identified Device or Use Problem1
- Ejection Problem1
- Failure to Advance1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Capsular Bag Tear1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2026Failure to Advance; Scratched Material
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50532542 ↗MalfunctionMar 2, 2026Mechanical Jam; Ejection Problem
IT WAS REPORTED THAT A PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) GOT STUCK AT THE TIP OF THE CARTRIDGE. THE LENS WAS PARTIALLY INSERTED. IT WAS ALSO REPORTED THE LENS WAS INSERTED IN THE PATIENT'S RIGHT EYE AND REMOVED DURING THE SAME PROCEDURE. THE PATIENT FULLY RECOVERED. THE DIRECTIONS FOR USE WERE FOLLOWED, AND THERE WAS A DELAY IN THE PROCEDURE, BUT NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED OUTSIDE OF STANDARD CARE. THERE WERE NO VITRECTOMY, INCISION ENLARGEMENT, OR SUTURES REQUIRED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24492198 ↗MalfunctionDec 28, 2023Break; Mechanical Problem; Break; Mechanical Problem
IT WAS REPORTED THAT A PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) PLUNGER STARTED TO OVERRIDE THE LENS. THERE IS NO EVIDENCE OF DAMAGE ON THIS CARTRIDGE. THE NURSE WAS ABLE TO SUCCESSFULLY RETRIEVE THE LENS, LOAD IT MANUALLY IN A SEPARATE CARTRIDGE, AND USED A PLATINUM INSERTER. THE LENS WAS DEFECTIVE TO SOME DEGREE. THERE WAS PATIENT CONTACT AS THE CARTRIDGE TIP CONTACTED THE EYE. AN INCISION ENLARGEMENT, SUTURE, OR VITRECTOMY WAS NOT REQUIRED. THERE WAS NO NEED FOR ANY FURTHER PATIENT INTERVENTION. THE PATIENT WAS DISCHARGED HOME WITHOUT ANY ISSUE AND HAS FULLY RECOVERED. THROUGH FOLLOW UP IT WAS CONFIRMED THERE WAS NO DAMAGE TO THE CARTRIDGE AT ALL. IT WAS REPORTED THE CARTRIDGE WAS LUBRICATED WITH A COMPETITOR BALANCE SALT SOLUTION (BSS) WITH ADDED ADDITIVES. 0.5 MILLIMETER (ML) OF EPINEPHRINE WAS ADDED TO THE BSS BAG (500 ML). THIS SOLUTION IS USED TO PRIME THE CARTRIDGE WITH. THE LENS IS NOT AVAILABLE AS IT WAS LOST IN THE COMMOTION DURING THE SURGERY, HOWEVER, THE INSERTER WITH CARTRIDGE IS AVAILABLE FOR RETURN. NO FURTHER INFORMATION TO PROVIDED. .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18415857 ↗MalfunctionDec 28, 2023Break; Mechanical Problem; Mechanical Jam; Break; Mechanical Problem; Mechanical Jam
IT WAS REPORTED THAT A PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) PLUNGER STARTED TO OVERRIDE THE LENS AND WOULD NO LONGER PUSH THE LENS FORWARD. THE PLUNGER THEN EXITED THE END OF THE CARTRIDGE WITH THE LENS STILL PARTIALLY STUCK INSIDE THE CARTRIDGE. THE SIDE OF THE CARTRIDGE HAS BEEN DAMAGED BY THE PLUNGER. AND THE LENS WAS DEFECTIVE TO SOME DEGREE. THERE WAS PATIENT CONTACT AS THE CARTRIDGE TIP CONTACTED THE EYE. AN INCISION ENLARGEMENT, SUTURE, OR VITRECTOMY WAS NOT REQUIRED. THERE WAS NO NEED FOR ANY FURTHER PATIENT INTERVENTION. THE PATIENT WAS DISCHARGED HOME WITHOUT ANY ISSUE AND HAS FULLY RECOVERED. THROUGH FOLLOW UP IT WAS LEARNED THAT THE TIP OF THE CARTRIDGE WAS NOT DAMAGED. THERE WAS HOWEVER A LONG OUTWARD PROTRUSION ALONGSIDE THE CARTRIDGE AS IF THE PLUNGER HAD SCRAPED THE CARTRIDGE AS IT WAS BEING MOVED. IT WAS REPORTED THE CARTRIDGE WAS LUBRICATED WITH A COMPETITOR BALANCE SALT SOLUTION (BSS) WITH ADDED ADDITIVES. 0.5 MILLIMETER (ML) OF EPINEPHRINE WAS ADDED TO THE BSS BAG (500 ML). THIS SOLUTION IS USED TO PRIME THE CARTRIDGE WITH. THE LENS IS NOT AVAILABLE AS IT WAS LOST IN THE COMMOTION DURING THE SURGERY, HOWEVER, THE INSERTER WITH CARTRIDGE IS AVAILABLE FOR RE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18415338 ↗InjuryMay 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS (IOL) WAS INSERTED INTO PATIENT'S OPERATIVE EYE, HOWEVER, WAS REMOVED DUE TO CAPSULE TEAR. THE INCISION WAS ENLARGED AND A 3 PIECE LENS WAS USED AS THE REPLACEMENT. A VITRECTOMY WAS NOT REQUIRED. THERE WAS NO PATIENT HARM. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Capsular Bag Tear
MAUDE report 14505646 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page