Skip to content

TECNIS SIMPLICITY DIB00 (REF DIB00U0135)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TECNIS SIMPLICITY TECNIS EYHANCE U 13.5D · Class 3 · Implant

All models of TECNIS SIMPLICITY (112)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 05050474731622), received between Jul 30, 2021 and Aug 16, 2026 — classified as 7 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem6
  • Break5
  • Contamination1
  • Failure to Advance1
  • Failure to Fold1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • Visual Disturbances3
  • Blurred Vision2
  • Halo2
  • Capsular Bag Tear1
  • Endophthalmitis1
  • Eye Irritation1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 16, 2026Contamination

    It was reported that, during the initial stages of implanting a single-piece preloaded monofocal intraocular lens (IOL), a clear plastic particle was observed in the anterior chamber. The lens was not implanted, and the particle was retrieved. The reporter indicated that the particle was not believed to have originated from the IOL. No patient complications or adverse effects were reported. A backup lens was used to complete the procedure. No further information was provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50267130 ↗
  • InjuryJan 12, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER THE PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE, THE PATIENT WAS UNABLE TO ADAPT TO THE LENS AND NEAR TARGET. THE LENS WAS SUBSEQUENTLY EXPLANTED DURING A SECONDARY SURGERY AND REPLACED WITH ANOTHER JOHNSON AND JOHNSON IOL OF THE SAME MODEL WITH +10.5 DIOPTER. THE ISSUE WAS FIRST OBSERVED DURING A POST-OPERATIVE EXAMINATION AND PERSISTED FOR APPROXIMATELY TWO MONTHS AND THREE WEEKS. NO UNPLANNED SURGICAL INTERVENTIONS SUCH AS, INCISION ENLARGEMENT, SUTURES, OR VITRECTOMY WERE REQUIRED. IT WAS UNKNOWN IF THERE WERE ANY SURGICAL DELAYS OR MEDICATION PRESCRIBED THAT WAS OUTSIDE OF THE STANDARD OF CARE. DIRECTIONS FOR USE WAS FOLLOWED. THE PATIENT HAS FULLY RECOVERED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Visual Disturbances

    MAUDE report 24032866 ↗
  • MalfunctionMay 12, 2025Break; Break

    DURING THE LOADING PROCEDURE, A CRACK WAS NOTED IN THE PERIPHERAL OF THE PRE-LOADED DIB00 MODEL INTRAOCULAR LENS (IOL). NO ACTION WAS NEEDED, AND THERE WAS NO PATIENT HARM. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22005481 ↗
  • InjuryFeb 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT PHACOEMULSIFICATION SURGERY FOR THE RIGHT EYE, DURING WHICH A PRELOADED INTRAOCULAR LENS (IOL) WAS IMPLANTED. POST-INSERTION OF THE LENS, THERE WAS A NOTED VITREOUS LOSS, LEADING TO THE EXPLANTATION OF THE LENS AND PERFORMANCE OF A VITRECTOMY. SUBSEQUENTLY A NON JNJ LENS WAS IMPLANTED INCLUDING INCISION ENLARGEMENT AND SUTURES PERFORMED. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Vitreous Loss

    MAUDE report 21329311 ↗
  • MalfunctionNov 21, 2023Break; Failure to Advance; Break; Failure to Advance

    IT WAS REPORTED THAT THE PRELOADED MONOFOCAL TECNIS SIMPLICITY INTRAOCULAR LENS (IOL) WAS REMOVED AND REPLACED FROM THE PATIENT'S RIGHT EYE DURING THE SAME PROCEDURE DUE TO THE CRACKED OPTICS OF THE LENS. SIGNIFICANT RESISTANCE WAS ENCOUNTERED UPON ROTATION/INJECTION. THERE WAS NO INCISION ENLARGEMENT, SUTURES, OR VITRECTOMY NEEDED. THE LENS WAS REPLACED WITH ANOTHER LENS, BUT WITH AN UNKNOWN DIOPTER. THE PATIENT'S POST-PROCEDURE STATUS IS FINE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18187459 ↗
  • MalfunctionOct 5, 2023Break; Break

    IT WAS REPORTED THE JOHNSON AND JOHNSON(JNJ) INTRAOCULAR LENS (IOL) WAS WASTED. THE CUSTOMER NOTICED THE LENS WAS BROKEN BEFORE IT WAS USED IN THE PATIENT¿S RIGHT EYE. THE LENS WAS WASTED. THERE WAS NO MEDICAL OR SURGICAL INTERVENTION SUCH AS INCISION ENLARGEMENT, VITRECTOMY, OR SUTURES. REPORTEDLY, THE PATIENT HAS PRE-EXISTING BLEPHAROPHIMOSI. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17882702 ↗
  • InjurySep 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PRELOADED JOHNSON AND JOHNSON (JNJ) INTRAOCULAR LENS (IOL) WAS IMPLANTED INTO THE PATIENT¿S RIGHT EYE. THE PATIENT EXPERIENCED AN UNUSUAL INFLAMMATION CENTERED AROUND THE IOL. THEIR VISUAL ACUITY DECREASED, AND THEY EXPERIENCED PHOTOPSIAS AND PHOTOPHOBIA. THE EYE WAS CULTURED AND A TEAR ORGANISM WAS NOT SEEN. THE PATIENT WAS ADMINISTERED INTRAVITREAL INJECTION OF VANCOMYCIN, CEFTAZADIME, AND DEXAMETHASONE. IOL IS NOT AVAILABLE FOR RETURN AS IT REMAINS IMPLANTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Blurred Vision; Endophthalmitis; Eye Irritation; Eye Pain; Inflammation; Visual Disturbances

    MAUDE report 17733671 ↗
  • InjuryMar 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER IMPLANTING THE INTRAOCULAR LENS (IOL) INTO THE PATIENT¿S RIGHT EYE THEY EXPERIENCED FLOATERS AND RINGS WHICH WERE WORSENING. THE IMPLANT WAS A STANDARD PROCEDURE WITH NO COMPLICATIONS SUCH AS A CAPSULE TEAR, VITRECTOMY, OR SUTURES. THE POST OPERATIVE STATUS AFTER THE IOL IMPLANT WAS UNEVENTFUL. HOWEVER, THE SYMPTOMS WERE NOTICED ONE DAY POST-OPERATIVE AND PROGRESSIVELY WORSE AFTER WEEK ONE. TO AID THE PATIENT IN SEEING THE COMPUTER MONITOR THEIR MYOPIC TARGET WAS -0.41. HOWEVER, POST OPERATIVE DATA ON WEEK TWO (POW2) SHOWED WORSENING FLOATERS AND RINGS. THE PATIENT HAS PRE-EXISTING SYNERESIS WITH KNOWN POSTERIOR VITREOUS DETACHMENT (PVD) IN THEIR RIGHT EYE. HIGH MYOPE WITH LONG AXIAL LENGTH(AL) 26.58. POST-OPERATIVE BEST CORRECTED VISUAL ACUITY (BCVA) IS 20/30 NO HOLES OR TEARS, SYNERESIS PVD IS STABLE. WHEN EXAMINED POST-OPERATIVELY, THE PATIENT DESCRIBED PERFECT CIRCLES DIFFUSELY WITHIN HIS VISION ALL DIFFERENT SIZES WITH A CENTRAL PINPOINT DOT IN THE MIDDLE. THE CIRCLES HAVE GOTTEN WORSE SINCE THEIR POST EXAM ON WEEK TWO. PATIENT MOSTLY NOTICES THE CIRCLES WHEN OUTDOORS AND PLAYING TENNIS. LESS PROMINENT IN THE CLINIC WITH DILATION DROPS IN PLACE. PATI

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Flashers; Halo; Vitreous Floaters

    MAUDE report 16621449 ↗
  • MalfunctionJan 30, 2023Break; Failure to Fold; Break; Failure to Fold

    IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS (IOL) HAPTIC WAS FOLDED AND STICKING OUT OF THE FOLDER AND BENT. THERE WAS NO USE ERROR. THE EVENT WAS OBSERVED DURING HANDLING BUT PRIOR TO INSERTION INTO THE EYE. THERE WAS NO PATIENT CONTACT AND NO PATIENT INJURY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING A BACK-UP LENS. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16267690 ↗
  • MalfunctionSep 7, 2022Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT AFTER INSERTION INTO THE LEFT EYE (OS), IT WAS NOTICED BY THE SURGEON THAT THE HAPTIC WAS MISSING FROM THE INTRAOCULAR LENS (IOL). THE IOL WAS REMOVED AND REPLACED WITH ANOTHER JOHNSON & JOHNSON LENS, MODEL DIB00 13.5 DIOPTER. THERE WAS NO PATIENT INJURY AND NO ADDITIONAL SURGICAL INTERVENTION WAS REQUIRED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15375470 ↗
  • InjuryJun 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS INITIALLY REPORTED THAT A PATIENT EXPERIENCED HALOS/GLARE IN THEIR LEFT EYE POST A YTTRIUM ALUMINUM GARNET (YAG) PROCEDURE. SURGEON WAS NOT CERTAIN IF THERE WAS A CORRELATION BETWEEN THE NEWLY IMPLANTED INTRAOCULAR LENS (IOL) AND THE YAG PROCEDURE PERFORMED AFTER IMPLANTATION. ACCOUNT INDICATED THAT IT WAS NOT CLEAR IF THE YAG WAS PERFORMED PROACTIVELY AS PART OF THEIR PROCESS OR IF IT WAS DONE TO TREAT THE SYMPTOMS OF HALOS/GLARE. AT THAT TIME THE IMPLANT DATE AND IOL SERIAL NUMBER WERE NOT PROVIDED. THE LENS REMAINED IMPLANTED WITH NO PLANS TO EXCHANGE THE IOL. FURTHER INFORMATION RECEIVED ON 5/6/2022 PROVIDED THAT DURING THEIR LAST POST-OPERATIVE VISIT ON (B)(6) 2022 THE PATIENT WAS COMPLAINING OF BLURRY VISION, DIFFICULTY DRIVING AT NIGHT, GLARE/HALO AROUND LIGHTS, AND ISSUES WITH DEPTH PERCEPTION ESPECIALLY WITH INTERMEDIATE VISION. THE PATIENT INDICATED BEING TEMPORARILY IMPAIRED AND THE VISUAL ISSUES SIGNIFICANTLY AFFECT HER DAILY ACTIVITIES. ACCOUNT PROVIDED THAT THE PATIENT HAD THE IOL EYHANCE IMPLANTED IN (B)(6) 2021, THEN YAG WAS PERFORMED ON (B)(6) 2021 AS POSTERIOR CAPSULE OPACIFICATION (PCO) WAS OBSERVED. THE POST-OPERATIVE VISION WAS PROVIDED AS -20/25. SURGEO

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Halo; Visual Disturbances

    MAUDE report 14569828 ↗
  • InjuryMar 10, 2022Break; Break

    IT WAS REPORTED THAT UPON IMPLANTING THE INTRAOCULAR LENS (IOL), THE DOCTOR NOTICED A CRACK IN THE BODY OF THE LENS. THE IOL WAS REMOVED AND ANOTHER JOHNSON & JOHNSON LENS (SAME MODEL AND DIOPTER) WAS IMPLANTED AS A REPLACEMENT. THE INCISION WAS SLIGHTLY ENLARGED. THE IOL WAS DISCARDED AND WILL NOT BE RETURNED TO J&J. REPORTEDLY, THE PATIENT IS DOING WELL. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13727421 ↗
  • InjuryJul 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE SHOULDER OF THE HAPTIC OF THE PRELOADED INTRAOCULAR LENS (IOL) TORE THE POSTERIOR CAPSULE OF THE PATIENT'S LEFT EYE AND THE LENS WAS CUT OUT. ADDITIONAL INFORMATION RECEIVED CONFIRMED THAT NO ADDITIONAL MEDICAL/ SURGICAL INTERVENTIONS WERE REQUIRED OR ARE PLANNED. IT WAS ALSO CONFIRMED THAT THERE WAS NEITHER ANY DAMAGE TO THE HAPTIC OR LENS NOR ANY PRE-EXISTING PATIENT CONDITION THAT CONTRIBUTED TO THE EVENT. ANOTHER COMPETITOR LENS WAS PLACED AS THE REPLACEMENT AND THE PATIENT IS DOING FANTASTIC POST-OPERATION. NO FURTHER INFORMATION WAS AVAILABLE.

    Patient problems: Capsular Bag Tear

    MAUDE report 12256903 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page