TECNIS SIMPLICITY DIB00 (REF DIB00U0275)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TECNIS SIMPLICITY TECNIS EYHANCE U 27.5D · Class 3 · Implant
All models of TECNIS SIMPLICITY (112)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05050474731912), received between Aug 18, 2022 and Apr 22, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Ejection Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Ejection Problem2
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Difficult to Insert1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Capsular Bag Tear1
- Zonular Dehiscence1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2025Ejection Problem; Ejection Problem
IT WAS REPORTED THAT THE HAPTIC OF THE PRELOADED INTRAOCULAR LENS(IOL) WAS STUCK IN THE INJECTOR ALMOST HAD CUT THE LENS OUT. THE LENS WAS ABLE TO BE REMOVED WITHOUT ANY PATIENT HARM. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21884416 ↗MalfunctionFeb 7, 2025Ejection Problem; Ejection Problem
IT WAS REPORTED THAT TRAILING HAPTIC OF THE PRELOADED INTRAOCULAR LENS (IOL) WAS GETTING STUCK ON THE INJECTOR AND WAS ABLE TO BE RELEASED WITH A FORCEPS TO RIP THE HAPTIC OUT OF THE INJECTOR LEAVING THE HAPTIC OUTSIDE OF THE EYE. FROM ADDITIONAL INFORMATION IT WAS LEARNT THAT DURING IMPLANTATION OF THE LENS THE TRAILING HAPTIC OF THE LENS HAD STUCK IN THE PRELOADED CARTRIDGE. THE SURGEON HAD TO PULL THE LENS OUT WITH A FORCEPS AND IMPLANT THE LENS. SO THE CARTRIDGE AND THE SHOOTER HAD TOUCHED THE PATIENT'S EYE. THE LENS IS STILL IMPLANTED IN THE PATIENT'S EYE AS THERE WAS NOTHING WRONG WITH IT. THE PATIENT IS DOING WELL. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21335086 ↗MalfunctionAug 8, 2023Break; Break
IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WAS INSERTED INTO THE PATIENT'S RIGHT EYE. UPON INSERTION, THE LENS HAD TO BE CUT OUT AND REPLACED. THE DOCTOR NOTICED THAT THE HAPTIC WAS BROKEN ONCE THE PRE-LOADED LENS WAS PLACED IN THE EYE. A BACK UP LENS WAS UTILIZED WITH NO FURTHER COMPLICATIONS. IT WAS NOTED THAT THE BACK-UP LENS USED HAS THE SAME MODEL AND DIOPTER SIZE AS THE SUSPECT IOL. THERE WAS NO INCISION ENLARGEMENT, NO VITRECTOMY, NO SUTURES DONE. NO PATIENT POST-OP INJURY. THE SUSPECT IOL WAS DISCARDED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17481665 ↗InjuryAug 18, 2022Difficult to Insert; Adverse Event Without Identified Device or Use Problem; Difficult to Insert; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CUSTOMER PUT THE INTRAOCULAR LENS (IOL)IN THE PATIENT'S LEFT EYE BUT PATIENT'S EYE WOULD NOT HOLD IT. THEY HAD TO CUT THE IOL IN HALF TO TAKE IT OUT AND REPLACED IT WITH A NON JOHNSON AND JOHNSON IOL. IT WAS LEARNED THAT DURING THE PROCEDURE THERE WAS A CAPSULAR WEAKNESS WHICH LEAD TO A POSTERIOR CAPSULAR TEAR. HOWEVER, THE CUSTOMER DID NOT THINK THAT THERE WAS A PRODUCT QUALITY ISSUE. THERE WAS NO INCISION ENLARGEMENT, NO VITRECTOMY OR SUTURES DONE. THE PATIENT LEFT WITHOUT PROBLEMS. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Capsular Bag Tear; Zonular Dehiscence
MAUDE report 15255775 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page