Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05050474746626), received between Aug 6, 2024 and Aug 6, 2024 — classified as 1 injury.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Visual Impairment1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PRELOADED JOHNSON AND JOHNSON (JNJ) INTRAOCULAR LENS (IOL) WAS IMPLANTED IN THE PATIENT'S LEFT EYE. THERE WERE NO COMPLICATIONS SUCH AS A CAPSULE TEAR, VITRECTOMY, OR SUTURES. THE OUTCOME WAS HYPEROPIC SURPRISE. THE IOL IN THE PATIENT'S LEFT EYE WAS IMPLANTED AT 80 DEGREES AND FOUND TO BE AT 80 DEGREES AS INTENDED. UNCORRECTED VISUAL ACUITY (UCVA) IS 20/100 AND BEST CORRECTED VISUAL ACUITY (BCVA) 20/20. THE PATIENT'S MANIFEST REFRACTION (MRX) IN THEIR LEFT EYE WAS +1.00 SPHERICAL SO THE ASTIGMATISM WAS CORRECTED AS EXPECTED. THE MANIFEST REFRACTION SHOWS NO CYLINDER AS INTENDED. THE PATIENT WAS NOT OK WITH THEIR HYPEROPIC SURPRISE IN THEIR LEFT EYE. THEY ALSO FOUND THE SYMPTOM DEBILITATING AS THEY USED TO DRIVE BUT COULD NO LONGER DRIVE. CONSEQUENTLY, THE IOL WAS EXPLANTED AND REPLACED WITH THE SAME MODEL AND DIFFERENT DIOPTER, 25.5 DIOPTER (SERIAL (B)(6) DIU150 +25.5D). THE IOL IS IN THE BAG AND CENTERED. THE PATIENT IS HAPPY WITH THE OUTCOME. THEY ARE DOING WELL AND HEALING APPROPRIATELY. THE PATIENT'S RIGHT EYE HAD MYOPIC SURPRISE BUT THE PATIENT IS OK WITH IT AND IT'S NOT DEBILITATING. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Visual Impairment
MAUDE report 19925349 ↗
Other LENS, INTRAOCULAR, TORIC OPTICS devices
- TECNIS IOL ZCT150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS1 recall7 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS43 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS36 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS32 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS30 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS29 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS21 reports
- TECNIS IOL DIU150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page