TECNIS IOL DIU150 (REF DIU150U255)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TEC EYHANCE TOR II SMPLCTY 25.5D CYL1.50 · Class 3 · Implant

All models of TECNIS IOL (1344)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05050474746657), received between Jan 26, 2023 and Dec 10, 2025 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Inaccurate Delivery” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inaccurate Delivery3
  • Break2
  • Adverse Event Without Identified Device or Use Problem1
  • Application Program Problem: Dose Calculation Error1
  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Eye Injury2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Discomfort1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 10, 2025Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    IT WAS REPORTED BY THE CUSTOMER THAT THE PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL), MODEL DIU150 WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD). INCORRECT DIOPTER SELECTED, RESULTING IN POOR DISTANCE VISION AND PATIENT DISSATISFACTION. LOCATION OF IMPLANT FOR THE OD AT 20 DEGREES. LENS WAS LATER EXPLANTED DURING A SECONDARY SURGICAL PROCEDURE. THE PATIENT OUTCOME IS UNKNOWN/NOT PROVIDED. THE EXPLANTED LENS WAS NOT KEPT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Discomfort

    MAUDE report 23765064 ↗
  • InjuryDec 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PRE-LOADED DIU150 MODEL INTRAOCULAR LENS (IOL) WAS REPORTED TO HAVE BEEN IMPLANTED IN THE PATIENT'S EYE. THE IOL WAS EXPLANTED AND REPLACED WITH A DIB00 23.0 DIOPTER IOL THAT WAS IMPLANTED IN THE PATIENT¿S OCULAR DEXTER (RIGHT EYE). PATIENT VISUAL ACUITY POST-LENS EXCHANGE WAS REPORTED AS 20/40. THE SUSPECT IOL IS NOT AVAILABLE FOR RETURN AS IT WAS DISCARDED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Eye Injury

    MAUDE report 20873482 ↗
  • MalfunctionOct 18, 2024Crack; Crack

    IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) CARTRIDGE BROKE DURING LENS INJECTION. THE SURGEON USED BACKUP LENSES AND THE CARTRIDGE TIP CAME INTO CONTACT WITH THE PATIENT. NO INJURIES OR MEDICAL INTERVENTIONS OCCURRED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20482383 ↗
  • MalfunctionJan 26, 2023Inaccurate Delivery; Inaccurate Delivery

    IT WAS REPORTED THAT THE SURGEON WAS ABLE TO MANIPULATE THE PRELOADED INTRAOCULAR LENS (IOL) HAPTICS WITH SURGICAL INSTRUMENTS TO ENSURE THE LENS WAS SAFELY IMPLANTED IN THE RIGHT EYE. THERE WAS NO DAMAGE OR ISSUE WITH THE LENS OR HAPTIC DURING INSERTION. THERE WAS NO ISSUE WITH THE LENS OR PATIENT AFTER THE LENS WAS IMPLANTED. THE PATIENT HAS NO COMPLICATIONS POST-OPERATIVELY AND HAS GOOD OUTCOMES. THERE ARE NO PLANNED INTERVENTIONS. THE LENS REMAINS IMPLANTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16247671 ↗
  • MalfunctionJan 26, 2023Break; Inaccurate Delivery; Break; Inaccurate Delivery

    IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS (IOL) TRAILING HAPTIC GOT STUCK IN THE CARTRIDGE, SCRUNCHING IT, WHILE BEING INJECTED IN THE RIGHT EYE. THE LENS WAS FULLY INSERTED AND THEN REMOVED AND REPLACED IN THE SAME PROCEDURE. NO ADDITIONAL SURGICAL MANEUVERS WERE REQUIRED. WHEN THE LENS WAS REMOVED THERE WAS MINOR INJURY TO THE IRIS, HOWEVER NO SURGICAL OR MEDICAL INTERVENTION WAS REQUIRED. ANOTHER JOHNSON AND JOHNSON IOL WAS IMPLANTED AS REPLACEMENT (SAME MODEL AND DIOPTER). THE INCISION WAS ENLARGED. THE LENS IS NOT AVAILABLE FOR RETURN AND EVALUATION. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Eye Injury

    MAUDE report 16247601 ↗
  • MalfunctionJan 26, 2023Break; Inaccurate Delivery; Break; Inaccurate Delivery

    IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS (IOL) TRAILING HAPTIC GOT STUCK IN THE CARTRIDGE, SCRUNCHING IT, WHILE BEING INJECTED IN THE RIGHT EYE. THE LENS WAS PARTIALLY INSERTED AND THEN REMOVED. NO ADDITIONAL SURGICAL MANEUVERS WERE REQUIRED. A BACK-UP LENS WAS USED (SAME MODEL AND DIOPTER). THE LENS IS NOT AVAILABLE FOR RETURN AND EVALUATION. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16247557 ↗

Other LENS, INTRAOCULAR, TORIC OPTICS devices

All LENS, INTRAOCULAR, TORIC OPTICS devices →