TECNIS IOL DIU225 (REF DIU225U210)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TEC EYHANCE TOR II SMPLCTY 21.0D CYL2.25 · Class 3 · Implant

All models of TECNIS IOL (1344)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05050474747159), received between Dec 8, 2022 and Mar 12, 2026 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Break2
  • Difficult to Insert1
  • Ejection Problem1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Visual Impairment2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Blurred Vision1
  • Capsular Bag Tear1
  • Headache1
  • Unspecified Eye / Vision Problem1
  • Visual Disturbances1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 12, 2026Break; Use of Device Problem; Ejection Problem

    IT WAS REPORTED THAT A SINGLE-PIECE PRELOADED TORIC INTRAOCULAR LENS (IOL) WAS INSERTED INTO THE PATIENT¿S EYE. DURING INSERTION, IT WAS OBSERVED THAT THE TRAILING HAPTIC BECAME ENTANGLED AROUND THE INSERTER. THE SURGEON ATTEMPTED TO DISENGAGE THE HAPTIC USING A TYER TOOL; HOWEVER, THE HAPTIC FRACTURED DURING THIS MANEUVER. THE AFFECTED IOL WAS EXPLANTED AND REPLACED WITH A BACKUP IOL OF THE SAME MODEL AND DIOPTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24581993 ↗
  • InjuryFeb 2, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PRELOADED JOHNSON AND JOHNSON (JNJ) INTRAOCULAR LENS (IOL) WAS IMPLANTED INTO THE PATIENT¿S EYE. THE PATIENT EXPERIENCED A MYOPIC SHIFT WHICH WAS IMPACTING THEIR DAILY ACTIVITIES. THE IOL WAS EXPLANTED AND REPLACED. THE REPLACEMENT LENS INFORMATION WAS NOT PROVIDED. THERE WERE NO COMPLICATIONS SUCH AS A CAPSULE TEAR, VITRECTOMY, OR SUTURES. THE PATIENT IS DOING WELL. NO FURTHER INFORMATION REGARDING THIS EVENT IS AVAILABLE.

    Patient problems: Visual Impairment

    MAUDE report 24229188 ↗
  • InjuryApr 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS (IOL) WAS EXPLANTED FROM THE PATIENT'S LEFT EYE AFTER TWO MONTHS OF IMPLANTATION. THE PATIENT REPORTED INTOLERABLE DYSPHOTOPSIA AND LOSS OF BEST CORRECTED VISION. IT WAS STATED THAT THE SYMPTOMS STARTED AS OF DAY ONE POST OP, BLURRED VISION, GHOSTING, BLUE TINT, VIBRATING EYE, HEADACHES AND HIS SIDE VISION HAD AN IRRITATING SHARP LIGHT. THESE SYMPTOMS DID NOT IMPROVE AT 1 WEEK OR AT FUTURE APPOINTMENTS. LENS WAS REPLACED WITH THE SAME MODEL AND SAME DIOPTER LENS IN A SECONDARY PROCEDURE. THERE WAS NO CALCULATION ISSUE. THERE WAS NO PATIENT INJURY. THERE WAS NO MEDICAL/SURGICAL INTERVENTION REQUIRED. PATIENT'S DAILY ACTIVITIES ARE NOT AFFECTED. THE PATIENT IS DOING GOOD. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: Blurred Vision; Headache; Unspecified Eye / Vision Problem; Visual Disturbances; Visual Impairment

    MAUDE report 19210371 ↗
  • MalfunctionFeb 16, 2023Break; Break

    IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) WAS DEFECTIVE. REPORTEDLY, IT¿S UNKNOWN IF THE IOL TOUCHED THE EYE. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16389183 ↗
  • InjuryDec 8, 2022Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS (IOL) WAS FULLY IMPLANTED AND REMOVED FROM THE RIGHT EYE BECAUSE A VITRECTOMY WAS NEEDED. PER THE DOCTOR, THERE WAS A POSTERIOR DEFECT NOTED IN THE CAPSULAR BAG AND DUE TO INSTABILITY OF THE LENS, THE DECISION WAS MADE TO EXPLANT. THE INSTABILITY OF THE LENS WAS DUE TO A CAPSULAR DEFECT. THERE WAS NO PRODUCT ISSUE. THE ISSUE WAS FIRST OBSERVED AFTER THE LENS WAS FULLY IMPLANTED AND THEN IT WAS CUT OUT AND REPLACED WITHIN THE SAME PROCEDURE. THE INCISION WAS ENLARGED AND ONE SUTURE AND A VITRECTOMY WERE REQUIRED. A NON-JOHNSON AND JOHNSON THREE-PIECE IOL WAS IMPLANTED AS REPLACEMENT. THE REPLACEMENT LENS WAS "DIALED IN" WITHOUT ISSUE. THE DOCTOR DOES NOT KNOW IF THE "DEFECT" WAS A RESULT OF THE ATTEMPT TO IMPLANT THE IOL OR IF IT WAS A PATIENT BIOLOGICAL ISSUE. THE PATIENT WAS DOING WELL AT THE POST-OP VISIT. NO FURTHER INFORMATION IS AVAILABLE WAS PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Capsular Bag Tear

    MAUDE report 15943491 ↗