TECNIS IOL DIU225 (REF DIU225U235)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TEC EYHANCE TOR II SMPLCTY 23.5D CYL2.25 · Class 3 · Implant
All models of TECNIS IOL (1344)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05050474747203), received between Nov 17, 2023 and Sep 3, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Contamination1
- Difficult to Insert1
- Ejection Problem1
- Material Fragmentation1
- Material Opacification1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Corneal Edema1
- Intraocular Pressure Increased1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT A PRELOADED MONOFOCAL TECNIS EYHANCE TORIC INTRAOCULAR LENS (IOL) WAS FULLY IMPLANTED, THEN REMOVED AND REPLACED DURING THE SAME PROCEDURE ON (B)(6) 2025, DUE TO A SHEARED-OFF HAPTIC. NO OTHER MEDICAL INTERVENTIONS WERE REPORTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT IS REPORTED TO BE IN GOOD CONDITION FOLLOWING THE PROCEDURE. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22957425 ↗MalfunctionApr 2, 2025Break; Ejection Problem; Break; Ejection Problem
IT WAS REPORTED THAT THE TRAILING HAPTIC OF THE PRELOADED INTRAOCULAR LENS (IOL) GOT STUCK IN THE CARTRIDGE DURING IMPLANTING THE LENS INTO THE PATIENT'S RIGHT EYE. IT WAS BEING HELD IN BY THE ROD. AFTER TRYING TO RETRACT AND ADVANCE SEVERAL TIMES, THE HAPTIC BROKE OFF. THE ISSUE WAS NOTICED DURING IMPLANTATION/APPLICATION. THE PART OF THE LENS IN THE EYE HAD TO BE CUT OFF AND REPLACED WITH ANOTHER LENS OF SAME MODEL AND DIOPTER IN THE SAME PROCEDURE. THERE WERE NO COMPLICATIONS. THERE WAS NO PATIENT INJURY. THERE WAS NO MEDICAL/SURGICAL INTERVENTION REQUIRED. FROM ADDITIONAL INFORMATION IT WAS LEARNT THAT THERE WAS ENOUGH OVD AS USUAL. SINCE TRAILING HAPTIC GOT STUCK IN CARTRIDGE WHILE THE RESTO OF THE IOL WAS NOT INSIDE THE EYE ON AC. THE ON AND OFF EFFORT TO TRY TO LIBERATE THE HAPTIC ENDED WITH THE BROKE THE TRAILING HAPTIC AT THE OPTIC JUNCTION. THERE WAS POST OP CORNEAL EDEMA. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Corneal Edema
MAUDE report 21756332 ↗InjurySep 6, 2024Difficult to Insert; Difficult to Insert
IT WAS REPORTED AFTER IMPLANTATION THE PRE-LOADED DIU225 MODEL INTRAOCULAR LENS (IOL) WAS PUSHED FROM EYE DUE TO HIGH INCREASED INTRAOCULAR PRESSURE (IOP). THE IOL WAS REMOVED AND REPLACED WITH A DIU225 23.5 DIOPTER IOL DURING THE SAME PROCEDURE. THERE WAS NO INCISION ENLARGEMENT, NO SUTURES, NO PROCEDURAL DELAYS, AND NO MEDICAL ATTENTION WAS REQUIRED HOWEVER, AN UNPLANNED VITRECTOMY WAS PERFORMED. IT IS UNKNOWN IF MEDICATION WAS PRESCRIBED. THE IOL DIRECTIONS FOR USE WERE FOLLOWED. PATIENT'S DAILY ACTIVITIES WERE NOT SIGNIFICANTLY AFFECTED. PATIENT STATUS WAS REPORTED AS FULLY RECOVERED. IT WAS INDICATED THE IOL ITSELF WAS DISCARDED HOWEVER, THE INJECTOR IS AVAILABLE FOR RETURN. PATIENT INFORMATION CANNOT BE PROVIDED DUE TO PERSONAL DATA PRIVACY LEGISLATION/POLICY. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Intraocular Pressure Increased
MAUDE report 20163579 ↗MalfunctionNov 17, 2023Contamination; Material Opacification; Contamination; Material Opacification
IT WAS REPORTED BY THE DOCTOR THAT HE IMPLANTED A TECNIS EYHANCE TORIC SIMPLICITY INTRAOCULAR LENS (IOL). ONCE THE LENS WAS IN THE EYE, HE NOTICED WHAT HE CALLED A SMALL PLAQUE ON THE OPTIC, (WHITE SPOT). HE WAS NOT SURE WHAT IT IS. THE LENS WAS LEFT IN THE EYE. HE FELT THAT IT WAS OUTSIDE OF THE PATIENT'S FIELD OF VISION. HE HAS A POST-OP VISIT ON NEXT DAY. ADDITIONAL INFORMATION RECEIVED STATED THAT THE DEFECT MUST HAVE BEEN THAT EMULSION ON THE LENS, BECAUSE IT¿S COMPLETELY GONE ON NEXT DAY DURING POST-OP EXAMINATION. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18161914 ↗
Other LENS, INTRAOCULAR, TORIC OPTICS devices
- TECNIS IOL ZCT150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS1 recall7 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS43 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS36 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS32 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS30 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS29 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS21 reports
- TECNIS IOL DIU150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page