Skip to content

TECNIS IOL DIU375 (REF DIU375U225)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TEC EYHANCE TOR II SMPLCTY 22.5D CYL3.75 · Class 3 · Implant

All models of TECNIS IOL (1344)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05050474748361), received between Jul 31, 2025 and Dec 8, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Intraocular Pressure Increased1
  • No Clinical Signs, Symptoms or Conditions1
  • Pigment Dispersion Syndrome1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PRELOADED TORIC INTRAOCULAR LENS (IOL) WAS EXPLANTED FROM THE PATIENT'S RIGHT EYE DUE TO "WRONG POWER". IT IS UNKNOWN IF THE PATIENT HAD A LOSS OF 2 OR MORE LINES OF BEST SPECTACLE CORRECTED VISUAL ACUITY (BSCVA). ANOTHER JOHNSON AND JOHNSON IOL WAS IMPLANTED AS REPLACEMENT (DIU375 21.0 DIOPTER). THERE WAS NO PATIENT INJURY. NO UNPLANNED SURGICAL INTERVENTION WAS REQUIRED. IT IS UNKNOWN IF MEDICATION OUTSIDE THE STANDARD OF CARE WAS PRESCRIBED. THERE ARE NO PLANNED SURGICAL INTERVENTIONS. THE PATIENT¿S OUTCOME INITIALLY REPORTED AS FINE; HOWEVER LATER IT WAS INDICATED TO BE UNKNOWN AND IT IS UNKNOWN IF THE REPLACEMENT LENS IMPROVED OR RESOLVED THE PATIENT'S SYMPTOMS. THERE IS NO INFORMATION AVAILABLE REGARDING WHETHER THE INCORRECT LENS POWER WAS DUE TO A CALCULATION ERROR OR A USE ERROR. IT WAS CONFIRMED THAT THE LENS DID NOT MALFUNCTION, ITS POWER MATCHED THE INTENDED SPECIFICATION, AND THE DEVICE DID NOT CAUSE OR CONTRIBUTE TO THE EVENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23753075 ↗
  • InjuryJul 31, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A MONOFOCAL TORIC INTRAOCULAR LENS IMPLANTED IN A LARGER CAPSULOTOMY, LEADING TO THE HAPTICS RUBBING AGAINST THE IRIS. THIS HAS RESULTED IN SIGNIFICANT PIGMENT DISPERSION AND ELEVATED INTRAOCULAR PRESSURE IN THE PATIENT. THE LENS REMAINS IMPLANTED IN THE PATIENT'S RIGHT EYE AND CANNOT BE REMOVED DUE TO THE POSITIONING OF THE HAPTICS WITHIN THE LARGE CAPSULOTOMY. THE TIMING OF THE EVENT, WHETHER DURING INSERTION OR POST-OPERATION, IS UNSPECIFIED. WHILE THE SYMPTOMS ARE DESCRIBED AS SIGNIFICANT, IT IS NOT CLEAR WHETHER THEY AFFECT THE PATIENT¿S DAILY ACTIVITIES. NO SURGICAL OR MEDICAL INTERVENTIONS HAVE BEEN UNDERTAKEN OR ARE PLANNED. THE PRODUCT REMAINS IMPLANTED AND WILL NOT BE RETURNED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Intraocular Pressure Increased; Pigment Dispersion Syndrome

    MAUDE report 22682883 ↗

Other LENS, INTRAOCULAR, TORIC OPTICS devices

All LENS, INTRAOCULAR, TORIC OPTICS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page