TECNIS IOL DFR00V (REF DFR00VU130)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TECNIS SYNERGY SIMPLICITY 13.0D · Class 3 · Implant
All models of TECNIS IOL (1344)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05050474750531), received between Dec 8, 2021 and Jan 13, 2023 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Scratched Material” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Scratched Material2
- Break1
- Crack1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Visual Disturbances2
- Blurred Vision1
- Flashers1
- Halo1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2023Scratched Material; Scratched Material
IT WAS REPORTED THE DFR00V WAS INSERTED INTO THE PATIENT'S OCULAR DEXTER (RIGHT EYE) WHEN A SCRATCH WAS NOTED BY THE SURGEON. THE IOL WAS PREPARED APPROPRIATELY BY THE TECHNICIAN IN THE ROOM WITH BALANCED SALT SOLUTION (BSS). IOL WAS REMOVED WITH MICRO SURGICAL TECHNOLOGY (MST) FORCEPS AND REPLACED WITH BACK UP LENS, WHICH WAS THE SAME MODEL AND DIOPTER SIZE. NO ADVERSE EFFECTS HAVE BEEN NOTED BY THE DOCTOR. THE DFR00V IOL IS NOT AVAILABLE FOR RETURN AS IT WAS DISCARDED. THROUGH FOLLOW-UP ADDITIONAL INFORMATION WAS RECEIVED CONFIRMING NO UNPLANNED INCISION ENLARGEMENT, NO SERIOUS PATIENT INJURY, NO SUTURE(S), AND NO VITRECTOMY. PATIENT OUTCOME POST-PROCEDURE WAS REPORTED AS: NO ADVERSE EFFECTS NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16162466 ↗MalfunctionOct 31, 2022Crack; Scratched Material; Crack; Scratched Material
IT WAS REPORTED THAT AFTER THE PRELOADED INTRAOCULAR LENS (IOL) WAS IMPLANTED IN THE RIGHT EYE, THE SURGEON NOTICED A SCRATCH. THE SURGEON DID NOT SEE A DEFECT IN THE LENS PRIOR TO INSERTION. THE SURGEON BELIEVES A PART OF THE INSERTER BROKE OFF AND SCRATCHED THE LENS. THE IOL WAS FULLY INSERTED AND THEN REMOVED AND REPLACED WITH ANOTHER JOHNSON & JOHNSON IOL (SAME MODEL AND DIOPTER). THERE WAS NO PATIENT INJURY. THROUGH FOLLOW-UP IT WAS LEARNED THAT THE TIP OF THE INSERTER WAS DAMAGED. IT WAS NOT COMPLETELY BROKEN OFF AND WAS STILL ATTACHED TO THE INSERTER LIKE A HANGNAIL. THE LENS WAS SCRATCHED IN TWO SEPARATE AREAS IN THE MIDDLE. WHEN THE SCRUB INSPECTED THE INSERTER WHEN LOADING, THERE WAS NO DEFECT. BEFORE INSERTING THE IOL, THE SURGEON DID NOT SEE A DEFECT UNDER THE MICROSCOPE. IT WAS AFTER SEEING THE SCRATCHES AND INSPECTING THE INSERTER AFTER INSERTION THAT THE WAS THE DEFECT NOTICED. NO OTHER SURGICAL/MEDICAL INTERVENTION WAS REQUIRED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15705362 ↗InjuryJun 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) WAS EXPLANTED FROM THE PATIENT'S OPERATIVE EYE DUE TO EXPERIENCING BLURRY VISION AND STAR BURSTS AT NIGHT. THERE WAS NO PATIENT INJURY REPORTED. ACCOUNT INDICATED THAT THE PATIENT'S VISUAL ACUITY 2 WEEKS POST SURGERY WAS 20/25. THE PATIENT WAS REPORTEDLY DOING GOOD AND CAN READ WITH OVER THE COUNTER READERS. THE EXPLANTED IOL IS NOT AVAILABLE TO BE RETURNED FOR EVALUATION. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Blurred Vision; Flashers
MAUDE report 14691794 ↗InjuryFeb 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WAS EXPLANTED FROM THE PATIENT'S LEFT EYE DUE TO PATIENT COMPLAINING OF DYSPHOTOPSIA. THE SUSPECT IOL WAS REPLACED IT WITH A DIB00 12.5 DIOPTER POWER. IT WAS NOTED THAT THERE WAS INCISION ENLARGEMENT. THERE WAS NO PATIENT INJURY REPORTED. THE EXPLANTED IOL IS NOT AVAILABLE FOR RETURN. AT 2-WEEK POST-OP IOL EXCHANGE PATIENT'S VISUAL ACUITY IS 20/25. THE PATIENT IS DOING WELL ENOUGH POST-EXPLANT, THAT THE DOCTOR RECOMMENDED CATARACT SURGERY ON FELLOW EYE. NO OTHER INFORMATION WILL BE AVAILABLE.
Patient problems: Visual Disturbances
MAUDE report 13421347 ↗MalfunctionJan 21, 2022Break; Break
IT WAS REPORTED THAT AFTER THE INTRAOCULAR LENS (IOL) WAS INSERTED IN THE RIGHT EYE, IT WAS NOTICED THAT IT WAS CRACKED. IT IS UNKNOWN IF THE HAPTIC OR OPTIC WAS CRACKED. THE IOL WAS REMOVED AND REPLACED IN THE SAME PROCEDURE. ANOTHER JOHNSON AND JOHNSON IOL WAS IMPLANTED (SAME MODEL AND DIOPTER). NO OTHER MEDICAL OR SURGICAL INTERVENTIONS WERE REQUIRED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13321308 ↗InjuryDec 8, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WAS EXPLANTED FROM THE PATIENT'S LEFT EYE DUE TO HALOS AND GLARES. A NON-JOHNSON & JOHNSON LENS WAS USED AS THE REPLACEMENT. THE INCISION WAS SLIGHTLY ENLARGED; HOWEVER, NO SUTURES WERE REQUIRED. THE PROCEDURE WAS COMPLETED AS PLANNED AND NO PATIENT INJURY WAS REPORTED. IT WAS INDICATED THAT THE DEVICE WAS DISCARDED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Halo; Visual Disturbances
MAUDE report 12965348 ↗
Other Lens, multifocal intraocular devices
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular144 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular1 recall66 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular113 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular112 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular110 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular107 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular104 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular93 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page