Skip to content

TECNIS IOL DFW375 (REF DFW375U090)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TECNIS SYNERGY TORII SMPCY 09.0D CYL3.75 · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05050474752818), received between May 10, 2023 and May 13, 2025 — classified as 2 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Blurred Vision1
  • Visual Disturbances1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAD POST OP NOCTURNAL DYSPHOTOPSIA AND HAD TROUBLE DRIVING AT NIGHT. THE PRELOADED TORIC INTRAOCULAR LENS, DFW375 9.0 DIOPTER WAS EXPLANTED AND REPLACED WITH DIU375 9.0 DIOPTER LENS IN A SECONDARY PROCEDURE. THERE WAS NO VITRECTOMY PERFORMED AND NO SUTURE WAS NEEDED. LENS IS NOT AVAILABLE FOR RETURN. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: Visual Disturbances

    MAUDE report 22014158 ↗
  • InjuryMay 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE DFW375 MODEL INTRAOCULAR LENS (IOL) WAS IMPLANTED INTO THE PATIENT¿S OCULAR DEXTER (RIGHT EYE). THE IOL WAS EXPLANTED IN A SECONDARY SURGICAL PROCEDURE DUE TO UNEXPECTED OUTCOMES OF BLURRY VISION AND REPLACED WITH A DFW225 8.5 IOL. THERE WAS NO PRODUCT QUALITY ISSUE THAT CONTRIBUTED TO IOL EXPLANT DECISION. THERE WAS NO INCISION ENLARGEMENT, NO SERIOUS PATIENT INJURY, NO SUTURE(S), AND NO VITRECTOMY. PATIENT OUTCOME POST-LENS EXCHANGE WAS REPORTED AS VISUAL ACUITY (VA) 20/20 DISTANCE AND 20/20 NEAR. IOL IS NOT AVAILABLE FOR RETURN. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Blurred Vision

    MAUDE report 16905176 ↗

Other Lens, multifocal intraocular devices

All Lens, multifocal intraocular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page