TECNIS ODYSSEY TORIC II IOL DRT150 (REF DRT1500170)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TEC ODYSSEY TOR II SMPCY 17.0D CYL1.50 · Class 3 · Implant
All models of TECNIS ODYSSEY TORIC II IOL (106)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05050474811294), received between Feb 21, 2025 and Sep 18, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Visual Impairment1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50822390 ↗MalfunctionNov 25, 2025Break; Break
IT WAS REPORTED THAT THE PRELOADED MULTIFOCAL TORIC INTRAOCULAR LENS (IOL) WAS FULLY INSERTED AND THEN REMOVED AND REPLACED IN THE SAME PROCEDURE DUE TO THE LENS BEING CRACKED. THE PROCEDURE WAS COMPLETED USING ANOTHER JOHNSON AND JOHNSON IOL OF THE SAME MODEL AND DIOPTER. NO MEDICAL INTERVENTION, SUCH AS VITRECTOMY, SUTURES, OR INCISION ENLARGEMENT, WAS REQUIRED. THERE WAS NO PATIENT INJURY. THE PATIENT IS DOING WELL POST-SURGERY. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23653113 ↗MalfunctionNov 25, 2025Break; Break
IT WAS REPORTED THAT THE PRELOADED MULTIFOCAL INTRAOCULAR LENS (IOL), MODEL DRT150, WAS FOUND TO BE DEFECTIVE AND SUBSEQUENTLY DISCARDED. THE EXTENT OF PATIENT CONTACT IS UNKNOWN/NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED. THE ISSUE WAS REPORTED WITH TWO LENSES. THIS REPORT CORRESPONDING TO THE PRODUCT (DRT150, SN (B)(6)). A SEPARATE REPORT WILL BE SUBMITTED FOR THE OTHER LENS (DRT150, SN (B)(6)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23653074 ↗MalfunctionJul 18, 2025Break; Break
IT WAS REPORTED THAT THE TORIC INTRAOCULAR LENS (IOL) WAS DELIVERED FROM THE CARTRIDGE WITH A CRACK FROM THE PERIPHERY GOING INWARD. IOL WAS CUT AND REMOVED FROM EYE. LOADING TECHNIQUE WAS APPROPRIATE, ADVANCING IOL IN FOLDED POSITION AND TUNNEL REPORTED NO RESISTANCE. VISCOELASTIC WAS USED AND THE DWELL TIME DURING THE DEVICE PREPARATION WAS APPROXIMATELY ONE MINUTE. NO MANIPULATION WAS REQUIRED TO DEPLOY THE DEVICE. SURGERY WAS COMPLETED IN SAME PROCEDURE. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22549898 ↗InjuryFeb 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A MULTIFOCAL INTRAOCULAR LENS (IOL) WAS EXPLANTED FROM A PATIENT'S RIGHT EYE BECAUSE OF SPECIFIC VISUAL ISSUES. THE PATIENT REPORTED THAT NEAR, INTERMEDIATE, AND NIGHT VISION WERE BLURRY AND FUZZY. A NON-JOHNSON & JOHNSON LENS WAS USED AS A REPLACEMENT (THE DIOPTER OF REPLACEMENT LENS WAS NOT PROVIDED). NO ADDITIONAL MEDICAL OR SURGICAL INTERVENTIONS SUCH AS VITRECTOMY, SUTURES, OR INCISION ENLARGEMENT WAS REQUIRED. THERE WAS NO PATIENT INJURY REPORTED. IT WAS INDICATED THAT THE PATIENT IS DOING BETTER, IMPROVED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Visual Impairment
MAUDE report 21434937 ↗
Other Lens, multifocal intraocular devices
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular144 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular1 recall66 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular113 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular112 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular110 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular107 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular104 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular93 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page