Skip to content

TECNIS ODYSSEY TORIC II IOL DRT150 (REF DRT1500255)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TEC ODYSSEY TOR II SMPCY 25.5D CYL1.50 · Class 3 · Implant

All models of TECNIS ODYSSEY TORIC II IOL (106)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05050474811461), received between Feb 13, 2025 and Jan 30, 2026 — classified as 3 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Halo1
  • No Clinical Signs, Symptoms or Conditions1
  • Visual Disturbances1
  • Visual Impairment1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 30, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT A PRELOADED MULTIFOCAL INTRAOCULAR LENS (IOL), MODEL DRT150, WAS IMPLANTED IN THE PATIENT¿S RIGHT EYE. THE LENS WAS LATER EXPLANTED DURING A SECONDARY SURGICAL PROCEDURE AND REPLACED WITH A DRN00V LENS (24.5 DIOPTER) DUE TO A MECHANICAL COMPLICATION. THE PATIENT OUTCOME WAS GOOD. BEST CORRECTED VISUAL ACUITY (BCVA) PREOPERATIVELY WAS 20/70-1, AND POSTOPERATIVELY WAS 20/60-1 WITH PINHOLE (PH) 20/40. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24223069 ↗
  • InjurySep 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DRT150 MODEL INTRAOCULAR LENS (IOL) WAS IMPLANTED IN THE PATIENT¿S OPERATIVE EYE. POST-OPERATIVELY, THE PATIENT HAD GLARE AND HALOS AT NIGHT DESPITE HAVING GOOD VISION, THUS THE IOL WAS EXPLANTED IN A SECONDARY SURGICAL PROCEDURE AND REPLACED WITH A DIU150 25.0 IOL. REPORTEDLY THERE WAS NO PATIENT ISSUES. PATIENT PROGNOSIS POST LENS EXCHANGE IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Halo; Visual Disturbances

    MAUDE report 23088958 ↗
  • InjuryFeb 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS(IOL) WAS EXPLANTED FROM PATIENT'S RIGHT EYE. AS THE PATIENT HAD A HYPEROPIC MISS; +1.66 D. PATIENT WAS UNABLE TO READ IN HER DAY TO DAY BASIS (BOOK, LABELS, SIGNS, ETC.) BECAUSE THE HYPEROPIC MISS. THE LENS WAS REPLACED WITH A DRT LENS WITH 26.0 DIOPTER IN A SECONDARY PROCEDURE. THERE WAS NO PATIENT INJURY. THERE WAS NO MEDICAL/SURGICAL INTERVENTION REQUIRED. AFTER NEW LENS THE PATIENT'S VISION WAS SLIGHTLY BLURRY BUT LATER IT IMPROVED AND IS DOING GOOD.

    Patient problems: Visual Impairment

    MAUDE report 21379421 ↗

Other Lens, multifocal intraocular devices

All Lens, multifocal intraocular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page