Skip to content

TECNIS ODYSSEY TORIC II IOL DRT225 (REF DRT2250140)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TEC ODYSSEY TOR II SMPCY 14.0D CYL2.25 · Class 3 · Implant

All models of TECNIS ODYSSEY TORIC II IOL (106)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05050474811829), received between Aug 11, 2025 and Aug 11, 2025 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Visual Impairment1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE DRT225 MODEL INTRAOCULAR LENS (IOL) WAS IMPLANTED IN THE PATIENT¿S OPERATIVE EYE. DUE TO COMPLAINTS OF ANISOMETROPIA, THE IOL WAS EXPLANTED IN A SECONDARY SURGICAL PROCEDURE AND REPLACED WITH AN ADVANCED TECHNOLOGY IOL (ATIOL). PATIENT PROGNOSIS POST LENS EXCHANGE IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Visual Impairment

    MAUDE report 22773363 ↗

Other Lens, multifocal intraocular devices

All Lens, multifocal intraocular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page