TECNIS ODYSSEY TORIC II IOL DRT375 (REF DRT3750220)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TEC ODYSSEY TOR II SMPCY 22.0D CYL3.75 · Class 3 · Implant
All models of TECNIS ODYSSEY TORIC II IOL (106)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05050474813168), received between Mar 20, 2025 and Feb 2, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Device Difficult to Setup or Prepare1
- Scratched Material1
- Use of Device Problem1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Blurred Vision1
- Vitreous Floaters1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 2, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PRELOADED JOHNSON AND JOHNSON (JNJ) INTRAOCULAR LENS (IOL) WAS IMPLANTED INTO THE PATIENT¿S LEFT EYE. DURING A POST-OPERATIVE EXAMINATION, THE PATIENT COMPLAINED OF THEIR NEAR VISION NOT BEING CLEAR AND THEIR VISION OVERALL IS FUZZY WITH SHADOWS. THE PATIENT HAD TROUBLE DRIVING AT NIGHT DUE TO HALOS. REPORTEDLY THE DIRECTIONS FOR USE WERE FOLLOWED. THE IOL WAS EXPLANTED AND REPLACED WITH A NON-JOHNSON AND JOHNSON LENS, MODEL ETN500 +21.5. THE PATIENT¿S BEST CORRECTED VISUAL ACUITY (BSCVA) PRE-OPERATIVE WAS 20/30 AND POST-OPERATIVE WAS 20/25 +1. THERE WAS NO INCISION ENLARGEMENT, VITRECTOMY, SUTURES, OR DELAY IN PROCEDURE. NO MEDICATION OUTSIDE THE STANDARD OF CARE. THE PATIENT HAS FULLY RECOVERED. NO FURTHER INFORMATION IS AVAILABLE REGARDING THIS EVENT.
Patient problems: Blurred Vision; Vitreous Floaters
MAUDE report 24229185 ↗InjuryMar 20, 2025Device Difficult to Setup or Prepare; Use of Device Problem; Scratched Material; Device Difficult to Setup or Prepare; Use of Device Problem; Scratched Material
WHEN SURGEON TRIED TO INSERT THE PRE-LOADED DRT375 MODEL INTRAOCULAR LENS (IOL), THE INSERTER DID NOT FIT INTO THE 2.4MM WOUND, AND THE INCISION WAS ENLARGED. THE DOCTOR THEN PROCEEDED TO TURN THE KNOB AND FELT HEAVY RESISTANCE. AFTER LENS INSERTION, MULTIPLE SCRATCHES WERE OBSERVED IN THE CENTER AND PERIPHERY OF THE LENS. ABOUT TWO (02) HOURS LATER THE IOL WAS EXPLANTED IN A SECONDARY SURGICAL PROCEDURE AND REPLACED WITH ANOTHER DRT375 22.0 DIOPTER IOL THAT WAS OBTAINED FROM ANOTHER FACILITY. THERE WAS NO SUTURES AND NO VITRECTOMY. PATIENT STATUS POST PROCEDURE WAS REPORTED AS FULLY RECOVERED. IT WAS REPORTED THE SURGICAL TECHNICIAN LOADING THE LENS HAS USED THIS INSERTER BEFORE. WHEN REVIEWING HE CONFIRMED THAT HE DID NOT OVERFILL INTO THE LENS CASE DURING HYDRATION. THE KNOB WAS TURNED MORE THAN HALF A ROTATION, PUSHING THE LENS TO THE TIP. SALES REP INFORMED HIM THAT HE NEEDS TO ONLY TURN A HALF ROTATION. BOTH BSS & OVD WERE USED. IT WAS INDICATED THE IOL DIRECTION FOR USE WERE NOT FOLLOWED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 21657475 ↗
Other Lens, multifocal intraocular devices
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular144 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular1 recall66 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular113 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular112 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular110 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular107 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular104 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular93 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page