Skip to content

VaccZyme™ Human Anti-Haemophilus Influenzae type b Enzyme Immunoassay Kit MK016Recalls & reported problems

The Binding Site Group Limited · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 12, 2018Class IITerminated

    A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.

    Z-1469-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other Antisera, All Types, H. Influenza devices

All Antisera, All Types, H. Influenza devices →