Skip to content

Optilite Freelite Lambda Kit LK018.OPT.ARecalls & reported problems

The Binding Site Group Limited · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05051700017800), received between Sep 15, 2026 and Sep 16, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “High Test Results” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K150658, K231290); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results2
  • Low Test Results2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 16, 2026High Test Results; Low Test Results

    The Binding Site were informed by the customer on (b)(6) 2026 that a falsely high Lambda Free Light Chain (FLC) and a falsely low Kappa Free Light Chain (FLC) patient results had been generated using the Freelite® Kappa Free Light Chain assay(LK016.OPT.A) and Freelite® Lambda Free Light Chain assay (LK018.OPT.A) on the Optilite® Analyser and released by the laboratory on (b)(6) 2026.The initial results were 2.3 mg/L for Kappa FLC and 19,722 mg/L for Lambda FLC. The discrepancy was identified when the results did not match the patient history. A redraw was tested on (b)(6) 2026 and produced results of 16 mg/L for Kappa FLC and 14.2 mg/L for Lambda FLC.Although no serious injury or change in patient management occurred, the incident is being reported as a precaution as the falsely high Lambda FLC result could potentially lead to serious injury if a similar issue were to recur and the result was considered in isolation. As stated in the IFUs (INS016.OPT.A and INS018.OPT.A), free light chain measurement results should always be interpreted in conjunction with other laboratory and clinical findings.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50792695 ↗
  • InjurySep 15, 2026High Test Results; Low Test Results

    The Binding Site were informed by the customer on (b)(6) 2026 that a falsely high Lambda Free Light Chain (FLC) and a falsely low Kappa Free Light Chain (FLC) patient results had been generated using the Freelite® Kappa Free Light Chain assay (LK016.OPT.A) and Freelite® Lambda Free Light Chain assay (LK018.OPT.A) on the Optilite® Analyser and released by the laboratory on (b)(6) 2026. The initial results were 2.3 mg/L for Kappa FLC and 19,722 mg/L for Lambda FLC. The discrepancy was identified when the results did not match the patient history. A redraw was tested on (b)(6) 2026 and produced results of 16 mg/L for Kappa FLC and 14.2 mg/L for Lambda FLC.Although no serious injury or change in patient management occurred, the incident is being reported as a precaution as the falsely high Lambda FLC result could potentially lead to serious injury if a similar issue were to recur and the result was considered in isolation. As stated in the IFUs (INS016.OPT.A and INS018.OPT.A), free light chain measurement results should always be interpreted in conjunction with other laboratory and clinical findings.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50772866 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K150658, K231290. One submission often covers many models, so check the product and lot details in each record.

  • Z-1043-2017 · Jul 5, 2016 · Terminated
    Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.

Other Lambda, Antigen, Antiserum, Control devices