Skip to content

CK-MB CAL CK2393Recalls & reported problems

Randox Laboratories Limited · The device is intended for use in the calibration of CK-MB assays in clinical chemistry systems. · Class 2

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Mar 25, 2022Terminated

    Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient Samples.

    Z-1097-2022 · Root cause: Under Investigation by firm · FDA record ↗

Other Calibrator, Secondary devices

All Calibrator, Secondary devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.