Skip to content

IMMUNOASSAY PREMIUM - TRI-LEVEL (IA PREMIUM 1, 2 AND 3) IA2633Recalls & reported problems

Randox Laboratories Limited · This product is intended for in vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes listed in the package insert. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 4, 2023Class IIIOpen, Classified

    Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range

    Z-2022-2023 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Other Multi-Analyte Controls, All Kinds (Assayed) devices

All Multi-Analyte Controls, All Kinds (Assayed) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page