Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jun 13, 2023Open, Classified

    Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.

    Z-2362-2023 · Root cause: Under Investigation by firm · FDA record ↗

Other Urease And Glutamic Dehydrogenase, Urea Nitrogen devices

All Urease And Glutamic Dehydrogenase, Urea Nitrogen devices →