Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05055343865224), received between Apr 23, 2025 and Nov 3, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Jam4
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 3, 2025Mechanical Jam; Mechanical Jam
TRINITY STD INTRODUCER / IMPACTOR HANDLE: STUCK TO TRINITY SHELL INTRA-OPERATIVELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23449954 ↗MalfunctionAug 14, 2025Mechanical Jam; Mechanical Jam
CORIN WERE INITIALLY INFORMED ON 15 MAY 2025 THAT DURING SURGERY THE TRINITY STD INTRODUCER / IMPACTOR HANDLE BECAME JAMMED WITH THE CUP DUE TO A LOOSE HANDLE. BASED ON THE INITIAL DETAILS PROVIDED, THIS EVENT WAS NOT CONSIDERED REPORTABLE. CORIN RECEIVED ADDITIONAL INFORMATION FROM THE REPORTER ON 23 JUL 2025 THAT DEEMED THIS EVENT REPORTABLE. IT WAS STATED THAT THE CUP HAD NOT REMAINED IMPLANTED AS ORIGINALLY THOUGHT. THE HANDLE AND CUP COULD NOT BE SEPARATED AFTER MULTIPLE ATTEMPTS AND THUS THE IMPACTED CUP WAS REMOVED FROM THE PATIENT'S ACETABULUM, AND AN ALTERNATIVE HANDLE AND CUP WERE LOCATED AND USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22803846 ↗MalfunctionJul 28, 2025Mechanical Jam; Mechanical Jam
DURING A TRINITY SURGERY 2 TRINITY STD INTRODUCER/IMPACTOR HANDLES COULD NOT UNSCREW FROM THE DEFINITIVE CUPS. THE IMPACTED CUPS HAD TO BE REMOVED AND ALTERNATIVE INSTRUMENTS/IMPLANTS LOCATED AND USED. THE REPORTER HAS STATED THAT THERE MAY HAVE BEEN WEAR TO THE THREADS OF THE HANDLES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22629362 ↗MalfunctionApr 23, 2025Mechanical Jam; Insufficient Device Problem Information; Mechanical Jam; Insufficient Device Problem Information
DURING TOTAL HIP PROCEDURE. AS THE SURGEON WAS IMPLANTING CUP THE SCREW IN THE CUP IMPACTOR POPPED OFF. NOT ALLOWING THE SURGEON TO REMOVE THE CUP. DELAYING CASE SINCE CUP WAS STUCK ON HANDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21890860 ↗
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page