Skip to content

Trinity Acetabular Hip System with HXLPE Liners 921.129Recalls & reported problems

Corin Ltd · STRAIGHT INTRODUCER/IMPACTOR HANDLE · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05055343865224), received between Apr 23, 2025 and Nov 3, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam4
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 3, 2025Mechanical Jam; Mechanical Jam

    TRINITY STD INTRODUCER / IMPACTOR HANDLE: STUCK TO TRINITY SHELL INTRA-OPERATIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23449954 ↗
  • MalfunctionAug 14, 2025Mechanical Jam; Mechanical Jam

    CORIN WERE INITIALLY INFORMED ON 15 MAY 2025 THAT DURING SURGERY THE TRINITY STD INTRODUCER / IMPACTOR HANDLE BECAME JAMMED WITH THE CUP DUE TO A LOOSE HANDLE. BASED ON THE INITIAL DETAILS PROVIDED, THIS EVENT WAS NOT CONSIDERED REPORTABLE. CORIN RECEIVED ADDITIONAL INFORMATION FROM THE REPORTER ON 23 JUL 2025 THAT DEEMED THIS EVENT REPORTABLE. IT WAS STATED THAT THE CUP HAD NOT REMAINED IMPLANTED AS ORIGINALLY THOUGHT. THE HANDLE AND CUP COULD NOT BE SEPARATED AFTER MULTIPLE ATTEMPTS AND THUS THE IMPACTED CUP WAS REMOVED FROM THE PATIENT'S ACETABULUM, AND AN ALTERNATIVE HANDLE AND CUP WERE LOCATED AND USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22803846 ↗
  • MalfunctionJul 28, 2025Mechanical Jam; Mechanical Jam

    DURING A TRINITY SURGERY 2 TRINITY STD INTRODUCER/IMPACTOR HANDLES COULD NOT UNSCREW FROM THE DEFINITIVE CUPS. THE IMPACTED CUPS HAD TO BE REMOVED AND ALTERNATIVE INSTRUMENTS/IMPLANTS LOCATED AND USED. THE REPORTER HAS STATED THAT THERE MAY HAVE BEEN WEAR TO THE THREADS OF THE HANDLES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22629362 ↗
  • MalfunctionApr 23, 2025Mechanical Jam; Insufficient Device Problem Information; Mechanical Jam; Insufficient Device Problem Information

    DURING TOTAL HIP PROCEDURE. AS THE SURGEON WAS IMPLANTING CUP THE SCREW IN THE CUP IMPACTOR POPPED OFF. NOT ALLOWING THE SURGEON TO REMOVE THE CUP. DELAYING CASE SINCE CUP WAS STUCK ON HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21890860 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources and updates