Skip to content

Unity Total Knee System 112.016.85Recalls & reported problems

Corin Ltd · Unity Tibial Insert PS Fixed Size 5 12mm · Class 2 · Implant

All models of Unity Total Knee System (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05055343875391), received between Jul 31, 2026 and Jul 31, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Unstable” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K113060); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unstable1

Patient problems

  • Joint Laxity1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 31, 2026Unstable

    Unity revision of the PS Insert after approx. 2 years due to instability.

    Patient problems: Joint Laxity

    MAUDE report 50098398 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113060. One submission often covers many models, so check the product and lot details in each record.

  • Z-0214-2024 · Oct 13, 2023 · Open, Classified
    Unity Total Knee System, Model Number 112.001.34. Primary and revision total knee arthroplasty
  • Z-0213-2024 · Oct 13, 2023 · Open, Classified
    Unity CR Femur Right, Size 6, Model Number 112.001.32 Primary and revision total knee arthroplasty
  • Z-0273-2023 · Oct 31, 2022 · Terminated
    Unity Total Knee System. Used for knee prosthesis in total knee replacement

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page