Aorfix AAA Stent Graft Body with Aorflex Delivery System SG-HBB-24-96-63-16-A12-22Recalls & reported problems
Lombard Medical Limited · Aorflex AAA Stent Graft Body with Aorflex Delivery System · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05055715606998), received between Nov 19, 2015 and Nov 19, 2015 — classified as 1 death.
The most frequently reported specific device problem is “Hole In Material” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Hole In Material1
Patient problems
- Aneurysm1
- Death1
- Rupture1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
DeathNov 19, 2015Hole In Material
EVENT 1: KINKED LEG DEVICE DURING INSERTION. EVENT 2: RUPTURED ANEURYSM / DEATH (RE-INTERVENTION). EVAR WAS PERFORMED ON (B)(6) 2015. THE CASE WAS A CHALLENGING, HIGH-ANGLE EVAR. BILATERAL CUTDOWNS WERE PERFORMED AND MAIN BODY AORFIX WAS INSERTED INTO THE LEFT ILIAC AND DEPLOYED AT THE RENAL ARTERIES. AFTER CANNULATING THE EVAR GRAFT FROM THE RIGHT SIDE, A COUPLE ATTEMPTS WERE MADE TO DELIVER AN AORFIX CONTRALATERAL LIMB. THE CONTRALATERAL LIMB WOULD NOT PASS AND WAS REMOVED FROM THE BODY. UPON REMOVAL, IT WAS NOTICED THAT THE AORFIX CONTRA LIMB HAD KINKED TRYING TO DELIVER IT. A GORE LIMB WAS THEN USED TO GRAFT THE RIGHT COMMON ILIAC FROM THE AORFIX FLOW DIVIDER DOWN TO THE ILIAC BIFURCATION AND A COOK LIMB EXTENSION WAS USED TO EXTEND FURTHER INTO THE LEFT COMMON ILIAC. EVERYTHING WAS THEN BALLOONED AND A FINAL ANGIO RUN WAS PERFORMED. NO ENDOLEAKS WERE REPORTED AT THAT TIME. PHYSICIAN CALLED AT 12:58AM ((B)(6) 2015) STATING THAT THE PATIENT'S ANEURYSM HAD RUPTURED. PATIENT HAD A CT CONFIRMING THE AORTIC RUPTURE AND WAS CURRENTLY IN THE HYBRID SUITE BEING TREATED. PHYSICIAN THEN SPOKE WITH (B)(6) TO DISCUSS NEXT TREATMENT STEPS. PATIENT OUTCOME IS UNKNOWN AT THIS TIME. UPDATE REC
Patient problems: Aneurysm; Death; Rupture
MAUDE report 5236416 ↗
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page