Aorfix AAA Stent Graft Body with Aorflex Delivery System SG-HBB-24-96-80-18-A12-22Recalls & reported problems
Lombard Medical Limited · Aorflex AAA Stent Graft Body with Aorflex Delivery System · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05055715607063), received between Aug 12, 2016 and Oct 27, 2016 — classified as 2 injuries.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Inadequacy of Device Shape and/or Size1
- Use of Device Problem1
Patient problems
- Aneurysm2
- Occlusion1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 27, 2016Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
ORIGINAL EVAR WAS PERFORMED ON (B)(6) 2015. A SMALL TYPE 4 ENDOLEAK REMAINED AT THE END OF THE PROCEDURE. A RE-INTERVENTION WAS PERFORMED ON (B)(6) 2016 TO TREAT THE PRESENCE OF A TYPE 2 ENDOLEAK DUE TO ANEURYSM ENLARGEMENT OF 1CM. AT THIS POINT, A TYPE 3B ENDOLEAK WAS IDENTIFIED BY THE CLINICIAN FROM A NEEDLE HOLE OF THE MAIN BODY WHERE THE STENT GRAFT WASN'T FULLY EXPANDED. THE PHYSICIAN CONSIDERED ADDING AN AORTIC CUFF INTO THE IMPLANTED AORFIX, HOWEVER, AS AN OPEN REPAIR HAD ALREADY BEEN INITIATED, THE AORFIX STENT GRAFT WAS EXPLANTED AND WAS REPLACED WITH A Y-GRAFT TO RESOLVE THE TYPE 3B ENDOLEAK.
Patient problems: Aneurysm
MAUDE report 6059975 ↗InjuryAug 12, 2016Use of Device Problem; Use of Device Problem
EVAR WAS PERFORMED (B)(6) 2016. EVENT: MAIN BODY AORFIX GRAFT PROXIMAL MIGRATION (OPEN REPAIR CONVERSION). CONTRALATERAL LIMB WAS INSERTED WITH SIGNIFICANT RESISTANCE FROM EXTERNAL ILIAC. LEFT IPSILATERAL LEG WAS CONSTRAINED 3CM PAST GATE. SURGEON ADVANCED THEN RELAXED. THIS WAS REPEATED WITH SOME SUCCESS. IT WAS NOTED THE CONTRALATERAL LUNDERQUIST WIRE AT THE BACK TABLE HAD GREATLY RETRACTED. SURGEON IMAGED PROXIMAL TO SEE WIRE POSITION AND NOTED FLOPPY END OF WIRE WAS NOW IN THE NOSECONE OF DELIVERY CATHETER. THIS RESULTED IN NOSECONE BUCKLING IN NECK ANGLE AND PUSHING GRAFT PROXIMALLY ABOVE THE LEFT RENAL ARTERY. POST ANGIO RUN OF DEPLOYED MAIN BODY AND CONTRALATERAL LEG, IMAGING REVEALED MAIN BODY GRAFT HAD MIGRATED ABOVE RENAL ARTERIES LEADING PHYSICIAN TO OPT FOR OPEN CONVERSION AND RESECTION OF IMPLANTED GRAFT(S).
Patient problems: Aneurysm; Occlusion
MAUDE report 5871152 ↗
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page