Aorfix AAA Stent Graft Body with Aorflex Delivery System SG-HBB-31-81-63-20-A12-22Recalls & reported problems

Lombard Medical Limited · Aorfix AAA Stent Graft Body with Aorflex Delivery System · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05055715611879), received between Jun 24, 2016 and Jun 24, 2016 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Aneurysm1
  • Occlusion1
  • Renal Failure1
  • Vascular Dissection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 24, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ORIGINAL EVAR WAS PERFORMED ON (B)(6) 2016. THE PROCEDURE WAS PERFORMED BY PLACING A MAIN BODY FROM THE RIGHT APPROACH FOLLOWING THE IFU. THE PHYSICIAN PLACED AND DEPLOYED A MAIN BODY (HBB-31-81), SO THAT THE FISHMOUTH TROUGH WOULD BE POSITIONED APPROXIMATELY 2 MM DISTALLY FROM THE LEFT LOWER RENAL ARTERY. THEREAFTER, HE RELEASED A SUPPORT TUBE, DEPLOYED AN IPSI-LATERAL LIMB, REMOVED THE DELIVERY SYSTEM FROM THE PATIENT'S BODY, AND INFLATED A BALLOON CATHETER FOR THE TOUCH-UP. IT WAS SEEMED TO THE PHYSICIAN THAT THE FISHMOUTH TROUGH WAS POSITIONED ABOUT 5 MM DISTALLY FROM THE LEFT LOWER RA AS A RESULT. THEN, HE PLACED A CONTRA-LATERAL LIMB (HBL-56-20) AND AN IPSI-LATERAL LIMB (HDE-20-51). AN ANGIOGRAPHY REVEALED A DEFINITIVE TYPE IA ENDOLEAK. BY INFLATING THE BALLOON CATHETER ADDITIONALLY, THE TYPE IA ENDOLEAK WAS DECREASED. THEREFORE, THE PHYSICIAN OBSERVED THE INSIDE OF THE MAIN BODY WITH AN ANGIOGRAPHY. NO ENDOLEAKS WERE CONFIRMED. THUS, HE CONCLUDED THAT IT WAS NOT A TYPE IV ENDOLEAK. THEREAFTER, HE PLACED AN EXTENDER (HPE-31-38) RIGHT BELOW THE RA. THE PROCEDURE WAS FINISHED AFTER COMPLETELY RESOLVING THE ENDOLEAK. A FOLLOW UP CT PERFORMED ON (B)(6) REVEALED THAT THE LEFT LOWE

    Patient problems: Aneurysm; Occlusion; Renal Failure; Vascular Dissection

    MAUDE report 5747548 ↗

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →