Aorfix AAA Stent Graft Plug In Leg with Aorflex Delivery System SG-HBL-106-12-A13-20Recalls & reported problems
Lombard Medical Limited · Aorfix AAA Stent Graft Plug In Leg with Aorflex Delivery System · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05055715612319), received between Jun 25, 2015 and Jun 25, 2015 — classified as 1 injury.
The most frequently reported specific device problem is “Patient-Device Incompatibility” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Patient-Device Incompatibility1
Patient problems
- Aneurysm1
- Claudication1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJun 25, 2015Patient-Device Incompatibility
THE ORIGINAL PROCEDURE WAS PERFORMED ON THE (B)(6) 2015. ON THE (B)(6) 2015, THE PHYSICIAN CONTACTED THE COMPANY INFORMING THE PATIENT PRESENTED WITH SUDDEN ONSET OF CLAUDICATION DUE TO OCCLUSION OF THE HIS LEFT ILIAC. CASE REVIEW: REVIEW OF THE PRE-OPERATIVE CT SHOWED THAT BOTH LEFT AND RIGHT ILIAC VESSELS WERE HIGHLY TORTUOUS. THE POST OPERATIVE IMAGE SHOWS THE STENT GRAFT DEPLOYED THROUGH THE LEFT COMMON ILIAC ANEURYSM AND WITH TWO EXTENSION PIECES (A 12MM DIAMETER LEG AND A 18MM DIAMETER DISTAL EXTENDER). A SHARP BEND, WHICH IS APPROXIMATELY 13MM X 5MM CAN BE SEEN ON THE PROXIMAL SIDE OF THE ILIAC ANEURYSM. A 51 X 18MM DISTAL EXTENDER HAS BEEN PLACED AT THE MOST DISTAL PART OF THE LIMB AND THE PROXIMAL END HAS A PRONOUNCED, NARROW FISHMOUTH WHERE IT HAS BEEN CONSTRAINED BY THE 12MM LIMB IN WHICH IT HAS BEEN PLACED. AN ALTERNATIVE TO USING AN OVERSIZED DISTAL EXTENDER IN THIS LOCATION WOULD HAVE BEEN TO HAVE USED A SHORT FLARED LIMB, SUCH AS A 56 X 18MM, SINCE IT HAS A 12MM DIAMETER PROXIMAL CONNECTION AND AN 18MM FINAL DIAMETER. THIS WOULD HAVE AVOIDED THE RISK OF THIS OCCLUSION. THE LOMBARD PROCTOR MET WITH THE CLINICIAN ON THE (B)(6) 2015, WHO TOLD THE PROCTOR THAT THE PATIEN
Patient problems: Aneurysm; Claudication
MAUDE report 4873967 ↗
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page