Maxi Sky 440 LE00000Recalls & reported problems
Arjohuntleigh AB · Maxi Sky 440 - Portable Ceiling Lift - Safe working load of 440 lb / 200 kg. · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05055982795319), received between Sep 26, 2022 and Dec 10, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Device Tipped Over1
Patient problems
- Fall1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 10, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE MAXI SKY 440 CEILING LIFT DETACHED AND FELL TOGETHER WITH THE PATIENT. NO INJURY WAS CLAIMED. THE MAXI SKY 440 PORTABLE CEILING LIFT MAY BE ATTACHED TO THE RAIL USING THE REACHER. THIS ACCESSORY IS COMPOSED OF THE REACHER ROD AND HOOK, CONNECTED TO EACH OTHER BY A MAGNET. TO CONNECT THE CEILING LIFT TO THE RAIL, THE REACHER'S HOOK NEEDS TO BE ATTACHED TO THE CEILING LIFT CARABINER AND CONNECTED TO THE RAIL TROLLEY. THE INSPECTION WAS PERFORMED BY ARJO TECHNICIAN IN PRESENCE OF DIRECTOR AND CAREGIVERS. IT REVEALED THAT THE REACHER'S HOOK WAS BENT. NO OTHER FAILURE WAS OBSERVED. THE LOAD TEST AT 200 KG WAS PERFORMED AND THE CEILING LIFT OPERATED CORRECTLY.
Patient problems: Fall
MAUDE report 20890936 ↗MalfunctionSep 26, 2022Device Tipped Over; Device Tipped Over
ARJO WAS INFORMED ABOUT THE INCIDENT WITH THE MAXI SKY 440 CEILING LIFT AND EASYTRACK FS INSTALLATION SYSTEM INVOLVEMENT. IT WAS REPORTED THAT THE CUSTOMER¿S STAFF TRANSFERRED A PATIENT FROM A BED TO A POWERED WHEELCHAIR. AFTER COMPLETING THE TRANSFER, THE STAFF LEFT THE PATIENT IN THE ROOM. WHEN THE PATIENT WANTED TO LEAVE THE ROOM IN THE WHEELCHAIR, THE EASYTRACK FS FRAME PARTIALLY FELL FORWARD TWISTING TO THE LEFT. THE PATIENT MANAGED TO DRIVE UNDER THE FRAME AND LEFT THE ROOM. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THE REPORTED EVENT. AFTER THE INCIDENT, THE INSTALLATION WAS INSPECTED BY AN ARJO SERVICE TECHNICIAN. THE SERVICE TECHNICIAN OBSERVED THAT THE LIFT ATTACHED TO THE TRACK MOVED FROM ONE END TO ANOTHER DUE TO FRAME COLLAPSE AND THAT THERE WERE WHITE MARKS ON THE INSTALLATION TRACK WHICH CAME AWAY WHEN WIPED. THE TECHNICIAN PLACED THE INSTALLATION IN THE CORRECT POSITION AND PERFORMED A FUNCTION TEST. NO PROBLEM OCCURRED AFTER THE REPAIR AND THE SYSTEM WAS PUT BACK INTO SERVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15486582 ↗
Other Lift, Patient, Non-Ac-Powered devices
- MEDLINE MDS88200DMedline Industries, Inc. · Lift, patient, non-ac-powered108 reports
- Medline MDS600ELBMedline Industries, Inc. · Lift, patient, non-ac-powered82 reports
- Likorall™ overhead lift 3123001Liko AB · Lift, patient, non-ac-powered1 recall1 death32 reports
- Medline MDS400ELBMedline Industries, Inc. · Lift, patient, non-ac-powered79 reports
- MEDLINE MDS88200HLMedline Industries, Inc. · Lift, patient, non-ac-powered77 reports
- Likorall™ overhead lift 3121001Liko AB · Lift, patient, non-ac-powered1 recall24 reports
- Likorall™ overhead lift 3124050Liko AB · Lift, patient, non-ac-powered1 recall18 reports
- Viking™ XL mobile lift 2040043Liko AB · Lift, patient, non-ac-powered1 recall14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page