Skip to content

Arjo Sling MAA4000M-XLRecalls & reported problems

Arjohuntleigh AB · Clip sling – padded legs - XL · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05055982796606), received between Dec 19, 2025 and Dec 19, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Abrasion1
  • Distress1
  • Fall1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THAT A RESIDENT FELL FROM AN ARJO SLING THAT WAS BEING USED WITH A NON-ARJO LIFT (JOYINCARE). DURING A VISIT TO THE CUSTOMER¿S FACILITY, IT WAS CLARIFIED THAT THE FALL WAS CAUSED BY THE DETACHMENT OF THE LEFT LEG CLIP FROM THE LIFT. IMMEDIATELY BEFORE IT DETACHED, AN UNRECOGNIZABLE SOUND (A SNAP) WAS HEARD BY THE CAREGIVER.THE CARE RECIPIENT WAS BEING GUIDED TO THE FLOOR AND SUSTAINED A SCRAPE ON THE HEAD; NO MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED. THE CAREGIVER REPORTED PSYCHOLOGICAL DISTRESS, INCLUDING SLEEP DISTURBANCE, AND PLANS TO SEE A TRAUMA THERAPIST.

    Patient problems: Abrasion; Distress; Fall

    MAUDE report 23847330 ↗

Other Lift, Patient, Non-Ac-Powered devices

All Lift, Patient, Non-Ac-Powered devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page