Skip to content

ALBAcyte Antibody Identification (16-Cell) Z473URecalls & reported problems

Alba Bioscience Limited · Reagent Red Blood Cells, ALBAcyte Antibody Identification (16-Cell) for tube techniques, 2-3% suspension.

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05060242471384), received between Dec 31, 2018 and Dec 12, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Inaccurate Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inaccurate Information1
  • Inadequate Instructions for Healthcare Professional1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2023

    AN END USER REPORTED A DAT POSITIVE RESULT WITH CELL 15 OF ALBACYTE REAGENT RED BLOOD CELLS FOR ANTIBODY IDENTIFICATION (16-CELL) PRODUCT Z473U, LOT V265925, EXPIRY 04DEC23, CELL 15 FOLLOWING A CONFIRMED. OFF LABEL USE OF THE DEVICE. THE END USER DID ORIGINALLY TEST BY THE CORRECT METHOD AS PER IFU AND GOT THE EXPECTED RESULT. NO RELEVANT TESTS/LABORATORY DATA RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18319074 ↗
  • MalfunctionDec 31, 2018Inadequate Instructions for Healthcare Professional; Inaccurate Information; Inadequate Instructions for Healthcare Professional; Inaccurate Information

    THE ANTIGEN PROFILE FOR US LICENSED REAGENT RED BLOOD CELLS ALBACYTE® ANTIBODY IDENTIFICATION CELLS 16-CELL (PRODUCT CODE Z473U, LOT NUMBER V194390) SUPPLIED TO CUSTOMERS CONTAINED AN ERROR. DURING IN-HOUSE TESTING OF THE US LICENSED REAGENT RED BLOOD CELL PRODUCTS ALBACYTE® ANTIBODY IDENTIFICATION CELLS 10-CELL NATIVE (PRODUCT CODE Z471U, LOT NUMBER V194387), 10-CELL PAPAIN-TREATED (PRODUCT CODE Z472U, LOT NUMBER V194389), ALBA BIOSCIENCE DISCOVERED AN ERROR IN THE ANTIGEN PROFILES FOR THESE PANELS. CELL # 9 WAS INCORRECTLY REPORTED AS JS(A+) WHILE THE DONOR IS ACTUALLY JS(A-). THE DEVIATION AFFECTS THE PRODUCT/LOT SUBJECT OF THIS REPORT BECAUSE IT SHARES THE SAME DONOR CELL WITH THE ABOVE PRODUCTS. THERE ARE NO ISSUES WITH ANY OTHER ASPECT OF THE PANEL. NO COMPLAINTS HAVE BEEN RECEIVED FROM THE FIELD IN RELATION TO THIS INACCURACY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8206171 ↗

Sources and updates