SOS FP-010Recalls & reported problems
Safe Obstetric Systems Limited · Fetal Pillow is a Single-use,sterile fetal head elevator consisting of a mechanism that elevates the fetal head to facilitate delivery during a Cesarean Section.Fetal Pillow is intended to elevate the fetal head and facilitate delivery of the fetus in women requiring a Cesarean Section at full dilation or those requiring a Cesarean Section after a failed instrumental vaginal delivery. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 05060388740009), received between Aug 23, 2023 and Aug 3, 2026 — classified as 7 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem7
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Hemorrhage/Blood Loss/Bleeding2
- Insufficient Information1
- Laceration(s)1
- Skull Fracture1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryAug 3, 2026Adverse Event Without Identified Device or Use Problem
It was reported by the Australian TAG that during a complicated birth, obstructed labour and fetal distress requiring manual fetal head disimpaction for planned emergency LUSCS, first attempted Fetal Pillow device and babe was born flat and suffered suspected subgaleal haemorrhage. Fetal pillow device recommended maximum inflation is 180cc and device found to have been inflated to 250cc. Baby required initial resuscitation and Special Care Nursery for observation, subsequently baby discharged home. Unknown long term effects.1216677-2026-00076 FP-010 Fetal Pillow 2026-07-0000460.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50117579 ↗InjuryApr 14, 2026Adverse Event Without Identified Device or Use Problem
PATIENT UNDERWENT CESAREAN DELIVERY FOR ARREST OF DESCENT. A FETAL PILLOW WAS FELT TO HAVE BEEN PROPERLY USED ACCORDING TO MANUFACTURER'S INSTRUCTIONS. NEWBORN WAS NOTED AT BIRTH TO HAVE A DEPRESSED SKULL FRACTURE WITH A SHAPE OF THE INDENTATION THAT COULD HAVE REFLECTED PRESSURE FROM THE FETAL PILLOW. IT COULD NOT BE DETERMINED WHETHER THE FRACTURE RESULTED FROM PRESSURE FROM THE FETAL PILLOW, FROM THE FORCES OF LABOR, OR FROM THE OPERATOR'S HAND THAT WAS USED TO ELEVATE THE HEAD FROM THE MATERNAL PELVIS AND THROUGH THE HYSTEROTOMY. 1216677-2026-00031 FP-010 FETAL PILLOW (B)(4).
Patient problems: Skull Fracture
MAUDE report 24872896 ↗InjuryDec 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CAESAREAN DELIVERY, THE BABY WAS OP (OCCIPUT POSTERIOR) PRESENTATION. THE DOCTOR WAS USING THE FETAL PILLOW FOR THE FIRST TIME; AND HE FOUND THE PILLOW WORKED WELL WITH A RESULTING SUCCESSFUL DELIVERY. IT WAS FOUND 2 TO 3 HOURS LATER THAT THE PATIENT WAS HEMORRHAGING. THE PATIENT WAS TAKEN BACK TO OR AND TWO LACERATIONS WERE NOTICED BEHIND THE UTERUS. THE PATIENT HAD TWO LACERATIONS NOTED SYMMETRICALLY ON THE POSTERIOR ASPECT OF THE BROAD LIGAMENT; THESE WERE ACTIVELY BLEEDING CAUSING THE HEMORRHAGE. THE PROVIDERS ARE CONCERNED THAT THIS MAY HAVE BEEN CAUSED BY THE FETAL PILLOW. POSSIBLY TEARS FROM STRETCHING. BLOOD TRANSFUSION WAS NEEDED DUE TO THE BLOOD LOSS. PATIENT ADMITTED TO THE ICU WITH SIX DAYS ADMITTANCE FOR RECOVERY. NO ADDITIONAL INFORMATION. FP-010 FETAL PILLOW 2025-11-0000507.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Laceration(s)
MAUDE report 23698188 ↗InjuryApr 29, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED FROM A CASE STUDY THAT AFTER AN UNCOMPLICATED PREGNANCY, A MALE BABY WAS DELIVERED BY EMERGENCY CAESAREAN SECTION AT 39 WEEKS OF GESTATION. BIRTH WAS COMPLICATED BY OBSTRUCTED LABOR, OCCIPUT POSTERIOR FETAL POSITION AND PROLONGED FETAL BRADYCARDIA. THE FETAL HEAD WAS IMPACTED AND, DESPITE THE USE OF FETAL PILLOW, THE HEAD WAS DIFFICULT TO EXTRACT FROM THE PELVIS. NEITHER FORCEPS NOR VACUUM DEVICES WERE USED TO FACILITATE DELIVERY. AT BIRTH, THE BABY WAS APNEIC, PALE AND BRADYCARDIC. THE BABY WAS INTUBATED AND HAD CHEST EXPANSION. CARDIAC COMPRESSIONS WERE ADMINISTERED AND THE BABY RECEIVED INTRAVENOUS ADRENALINE. SPONTANEOUS RESPIRATIONS AND CARDIAC OUTPUT WERE ESTABLISHED AT 10 MIN OF LIFE. THE BABY WAS ADMITTED TO THE NEONATAL INTENSIVE CARE UNIT FOR ONGOING CARE AND FOR THERAPEUTIC HYPOTHERMIA. TRUNCAL PETECHIAE WAS NOTICED AT 20 MINUTES OF AGE AND COAGULATION STUDIES WERE FOUND TO BE ABNORMAL. NO ADDITIONAL INFORMATION IS AVAILABLE. FP-010 FETAL PILLOW (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19201874 ↗InjurySep 22, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17802733 ↗InjurySep 22, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17802585 ↗InjuryAug 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 17605773 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page