Skip to content

DYNEX Agility 67000Recalls & reported problems

Dynex Technologies, Inc. · Automated ELISA Processing System with 12 Incubators · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jul 5, 2020Class IIOpen, Classified

    Control samples aspirated from wrong SmartKit on the Agility. Agility software was updated to v1.4.7 to resolve the issue. This leads to a risk that a control from another assay's SmartKit will be used instead of the correct control, which may lead to delayed patient results.

    Z-1222-2022 · Root cause: Software design · Recall details · FDA record ↗

Other Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use devices

All Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page