X80 / RADiant / PhotoElectric Therapy System RADiant AuraRecalls & reported problems
Xstrahl Limited · X80 / RADiant / PhotoElectric Therapy System · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05060494200084), received between May 19, 2024 and May 19, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 19, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON THE (B)(6) 2024, THE ARM JOINT BECAME LOOSE AND THIS CAUSED THE ARM TO DROP. THE RADIATION THERAPIST PUT HER SHOULDER IN THE PATH OF THE ARM TO STOP IT FROM DROOPING FURTHER. NO PATIENT INJURY. PRODUCT PROBLEM: SERVICE VISIT TO THE USER SITE FOUND THE FOLLOWING: MEASURED EXISTING UPPER ARM JOINT TORQUE WITH SHIMS INSTALLED (19.2NM) AS WELL AS LOWER JOINT (18.6NM). TIGHTENED JOINT SCREW TO APPROPRIATE 20NM SETTING. DURING THE TESTING AT THE USER SITE, IT WAS DISCOVERED THAT THE ARM JOINT BRAKE WOULD ENGAGE INTERMITTENTLY (APPROX 7 OF 10 TIMES). TESTED THE BASE UNIT PCB AND DISCOVERED FLICKERING OF LEDS WHEN FAILURE WOULD OCCUR AS IF CARD WAS BEING SHORTED OUT. ALSO NOTICED THE BASE HUB ROTARY BRAKES WERE NOT OPERABLE AT ALL (WOULD NOT DISENGAGE). DURING FAULT FINDING PROCEDURE THE BASE HUB BRAKE RESISTANCE READING WAS BELOW TOLERANCE AT APPROX 15.7 OHMS. CHECKED APPLICABLE BRAKE CABLING FOR DISCREPANCIES (DAMAGED OR CROSS WIRING) AND FOUND NONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19348479 ↗
Other System, Therapeutic, X-Ray devices
- Xstrahl Photoelectric Therapy SystemXstrahl Limited · System, Therapeutic, X-Ray1 recall0 reports
- INTRABEAM® Spherical Applicator 1.5 cm 304534-6000-557Carl Zeiss Meditec AG · System, Therapeutic, X-Ray1 recall0 reports
- INTRABEAM® Needle Applicator 304534-3001-200Carl Zeiss Meditec AG · System, Therapeutic, X-Ray1 report
- Balloon Applicator Kit 720342Icad, Inc. · System, Therapeutic, X-Ray1 death1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page