Skip to content

Procleix NAT Manager 1.2Recalls & reported problems

Biomedical Data Solutions Limited · Procleix NAT Manager Software is designed to collect data from Procleix assays, instruments, and pooling/lysis software, which allows operators to track individual donor blood samples and link them to the appropriate test results. The software collates the results of lysing, pooling and testing, assigns outcomes to individual samples according to a testing algorithm, and formats output files that can be sent to a Laboratory Information Management System (LIMS). Procleix NAT Manager Software also produces a variety of summary informational and analytical reports. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05060496770011), received between May 16, 2024 and May 16, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “False Negative Result” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 16, 2024False Negative Result; False Negative Result

    THE PROCLEIX PARVO/HAV ASSAY IS PERFORMING AS EXPECTED IN THIS EVENT. THIS REPORT PERTAINS TO THE MALFUNCTION OF THE PROCLEIX NAT MANAGER, THE MIDDLEWARE USED WITH THE PROCLEIX PARVO/HAV ASSAY ON PANTHER. BIOMEDICAL DATA SOLUTIONS LIMITED IS THE LEGAL MANUFACTURER OF THE PROCLEIX NAT MANAGER, AND GRIFOLS ACTS AS ITS DISTRIBUTOR. ON 16APR2024, GRIFOLS CUSTOMER (KREVNÍ CENTRUM ODBEROVÉ MÍSTO, CZECH REPUBLIC) REPORTED THAT A PARVO REACTIVE SAMPLE WAS RELEASED AS A NEGATIVE SAMPLE IN THE PROCLEIX NAT MANAGER. THE CUSTOMER DETECTED THAT THE SAMPLE WAS REACTIVE AND DID NOT RELEASE IT FOR TRANSFUSION. THE SAMPLE WAS CORRECTLY REPORTED AS '>100,000' FOR PARVO B19 BY PANTHER. HOWEVER, THE PANTHER/ PROCLEIX NAT MANAGER REPORTS FAILED TO MARK THE CEC (CONCENTRATION EQUAL TO OR EXCEEDED CUTOFF) FLAG FOR THIS SAMPLE, PANTHER DID NOT HIGHLIGHT THE SAMPLE IN THE RIGHT FORMAT, AND ALTHOUGH PROCLEIX NAT MANAGER CORRECTLY DISPLAYED THE REPORTED CONCENTRATION AS '>100,000', IT DISPLAYED THE MEASURED CONCENTRATION AS '-1' AND THE CONCLUSION AS 'BC' (BELOW CUTOFF). SEQUENCE OF EVENTS: ON PANTHER REPORT: PARVO IU/ML: >100, 000 FLAGS: NONE (MISSING CEC FLAG) NOT HIGHLIGHTED. ON PROCLEIX NAT MANAGER REPOR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19337926 ↗

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page