Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05391526210154), received between Oct 24, 2022 and Jun 4, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Inflation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inflation Problem1
  • Material Rupture1
  • Unstable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Material Rupture

    IT WAS REPORTED THAT THE BALLOON RUPTURED. THE TARGET LESION WAS LOCATION IN THE CORONARY ARTERY. A TRAPPER EXCHANGE DEVICE WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON BURST AT 18 ATM. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25402941 ↗
  • MalfunctionSep 3, 2024Inflation Problem; Inflation Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 09AUG2024. IT WAS REPORTED THAT DEVICE FAILED TO INFLATE. THE TARGET LESION WAS LOCATED IN THE LEFT CORONARY ARTERY. A TRAPPER WAS SELECTED FOR USE. DURING THE PROCEDURE, THE BALLOON WAS INFLATED AT 12 ATMOSPHERES HOWEVER BALLOON INFLATION DID NOT MEET EXPECTATIONS. THE DEVICE WAS REMOVED, AND PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT A LEAK WAS OBSERVED FROM A PINHOLE BURST ON THE SHAFT APPROXIMATELY AT 905MM FROM THE CATHETER TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20131808 ↗
  • MalfunctionOct 24, 2022Unstable

    INFLATED TO 16ATM ON DEVICE WHILE IN A 6FR GUIDE CATHETER. TRAPPER WOULD NOT HOLD WIRE IN PLACE WHILE ATTEMPTING TO REMOVE MICRO CATHETER.

    Patient problems: Insufficient Information

    MAUDE report 15657365 ↗