Skip to content

mo-Vis P002-77Recalls & reported problems

mo-Vis · All-Round Joystick R-net Heavy Duty · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jul 10, 2025Open, Classified

    Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

    Z-2459-2025 · Root cause: Device Design · FDA record ↗

Other Components, Wheelchair devices

All Components, Wheelchair devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page