SJM™ Masters SeriesHemashield™ 21CAVGJ-514 00Recalls & reported problems

St. Jude Medical, Inc. · St. Jude Medical Valved Graft Rotatable Mechanical Heart Valve · Class 3 · Implant

All models of SJM Masters SeriesHemashield (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05414734006026), received between Jul 30, 2020 and Jan 23, 2026 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Biocompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Biocompatibility1
  • Incomplete Coaptation1
  • Intravalvular regurgitation1
  • Perivalvular Leak1

Patient problems

  • Aortic Valve Insufficiency/ Regurgitation1
  • Hemorrhage/Blood Loss/Bleeding1
  • No Clinical Signs, Symptoms or Conditions1
  • Thrombus1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2026Incomplete Coaptation

    IT WAS REPORTED THAT ON (B)(6) 2025, A 21MM SJM MASTERS SERIES COATED AORTIC VALVED GRAFT WAS CHOSEN FOR A PROCEDURE. DURING PREPARATION, THE LEAFLET FUNCTION WAS TESTED USING A PLASTIC HOSE. DURING THE OPERATION, THE DOCTOR TRIED TO TEST THE VALVE LOBE AS NORMAL WHEN THE LEAFLETS STUCK. THE VALVE GOT REMOVED AND A SAME UNKNOWN VALVE WAS IMPLANTED WHILE PATIENT ON BYPASS. THERE WAS NO DELAY IN THE PROCEDURE. THE PATIENT WAS REPORTED STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24160964 ↗
  • InjuryJun 6, 2025Perivalvular Leak; Intravalvular regurgitation; Perivalvular Leak; Intravalvular regurgitation

    IT WAS REPORTED THAT ON (B)(6) 2025, A 21MM MASTERS SERIES HEART COATED AORTIC VALVED GRAFT WAS SUCCESSFULLY IMPLANTED IN A PATIENT. THE ANNULAR DIMENSIONS OF THE PATIENT'S NATIVE AORTIC VALVE WERE NOT GIVEN. THERE WAS NO DIFFICULTY EXPERIENCED ATTEMPTING TO IMPLANT THE 21MM MASTERS SERIES HEART COATED AORTIC VALVED GRAFT. POST-PROCEDURE, IT WAS NOTED THAT THE PATIENT HAD AORTIC REGURGITATION. THE DECISION WAS MADE EXPLANT THE 21MM MASTERS SERIES HEART COATED AORTIC VALVED GRAFT AND REPLACE IT WAS A 21MM REGENT HEART VALVE WITH FLEX CUFF AND NON-ABBOTT GRAFT DEVICE. THE PATIENT WAS REMOVED OFF CARDIOPULMONARY BYPASS (CPB) AND THEN PUT BACK ON CPB FOR FURTHER INTERVENTION. THERE WAS A CLINICALLY SIGNIFICANT DELAY IN PROCEDURE, BUT THE PATIENT WAS MEDICALLY MANAGED. THE PATIENT DID NOT HAVE ANY OTHER CLINICAL SIGNS OR SYMPTOMS AFTER THIS EVENT AND NO INITIAL OR PROLONGED HOSPITALIZATION. THE CAUSE OF THE AORTIC REGURGITATION IS RELATED TO A NOTABLE PERIVALVULAR LEAK OF THE 21MM MASTERS SERIES HEART COATED AORTIC VALVED GRAFT. THE PATIENT WAS STABLE AT THE TIME OF REPORT.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 22157950 ↗
  • InjurySep 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2023, A 21MM MASTERS SERIES AORTIC VALVED GRAFT WAS SELECTED FOR IMPLANT IN A PATIENT WITH A DEFORMED AND TORTUOUS AORTA. AFTER IMPLANTING THE VALVE, THE PATIENT WAS BLEEDING CONTINUOUSLY AND BECAME HEMODYNAMICALLY UNSTABLE REQUIRING A BLOOD TRANSFUSION. THE PHYSICIAN ALLEGED THAT THE DEFORMED AORTA WAS THE CAUSE OF THE BLEEDING POST IMPLANT OF THE VALVE. PATIENT STATUS IS UNKNOWN

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 17775719 ↗
  • InjuryJul 30, 2020Biocompatibility; Biocompatibility

    IN 2016, A 21MM MASTERS SERIES CAVGJ MECHANICAL VALVE WAS IMPLANTED. ON (B)(6) 2020, THE VALVE WAS EXPLANTED DUE TO THROMBUS THAT WAS UNABLE TO BE REMOVED VIA THROMBOEMBOLISATION. A COMPETITOR'S 23MM SORIN CONDUIT VALVE WAS SUCCESSFULLY IMPLANTED. THE PATIENT IS REPORTED TO BE IN STABLE CONDITION.

    Patient problems: Thrombus

    MAUDE report 10343951 ↗

Other HEART-VALVE, MECHANICAL devices

All HEART-VALVE, MECHANICAL devices →